AK 98 Dialysis Machine - Profiling
A:123
2)
Select the
UF rate
tab.
3)
Press
Profiling
.
4)
Press
UF rate
.
5)
Press
Mode
.
6)
Select graph method,
Linear
,
Step
, or
Interval
.
7)
Press
Confirm
.
8)
If the
profile
is
Step
or
Interval
:
Select
No of steps
or
No of intervals
,
whichever is applicable and adjust
according to the prescribed treatment.
9)
Set the
Start
value for UF rate. The
Stop
value for UF rate will be automatically
adjusted in relation to the
Start
value.
This is to keep the previously set UF
volume for the following treatment.
10)
Press
Activate
.
11)
Change
Na
+
(if applicable). See Section
“Profiling sodium without a preset
12)
Change
HCO
3
-
(if applicable). See Section
“Profiling bicarbonate without a
7.5.2
Profiling sodium without a preset model
Procedure
1)
Press the
Fluid
button.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...