ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 69 of 132
•
Press
ENTER
to save the edited value of the selection to
the transmitter.
Pressing
ESC
during relay output configuration editing
when the cursor is at the end of a line aborts parameter
editing without saving changes.
Pressing
ESC
from the Density, Nom Temp, Time Avg, or
Avg Time submenus when NOT actively editing a
parameter returns to the main Option menu. Pressing
ESC
again returns to the Data Display Screen.
The Option parameters may also be configured using the
ViscoSite software
–
refer to Section 5.8.
Network Configuration
4.6.3.1 Accessing the Network Configuration Setup
Submenu
•
From the Data Display, press the Option Setup button
on the keypad.
•
The Option submenu shown in Figure 4.10 is displayed.
•
Use the Up/Down arrow keys to scroll through the
Options entries to the Network Setup submenu.
The Network Setup submenu, used to configure the
ViscoSite transmitter for communication with a
remote
PC
across a Local Area Network (LAN), is shown in Figure 4.11.
Figure 4.11: Network Setup Sub Menu
If use of the ViscoSite software via LAN is desired, the
ViscoSite transmitter must be connected to a wired LAN via
the transmitter’s Ethernet port. The following parameters
must be properly configured for successful communication
with the transmitter over Ethernet:
IP
–
the assigned static IP address of the transmitter
GW
–
the IP address of the network gateway