
23
Figure 10: Alarms Panel
3.6 Inputs Panel
The Inputs panel shows the current status of the AccuLase-GPA™ IO board’s five analog inputs and four
digital inputs. See Figure 19.
Figure 11: Inputs Panel
By default, only two of the analog inputs are configured from the factory. Analog Input 1 is used for the
sample cell pressure transducer and is calibrated in units of torr. The typical pressure indicated here
should be approximately 250 torr. Analog input 5 is used for the sample cell temperature RTD input and
is calibrated in units of °C. The typical temperature indicated here should be 40°C. The other three analog
inputs can be configured as per the user’s requirements for inputs for any device (e.g. pressure transducer,
temperature sensor, etc.) that outputs data in a 4-20mA analog signal. Configuration of the analog inputs
is done from the AccuLase-GPA™ web GUI.