Revision 4
41
August 02, 2019
a)
Analog Output Full Scale
- These fields can be used to set the full scale range for the
H
2
S, SO
2
and COS, the Air Demand and whichever parameter is selected for Analog Out
4. If one of these fields is selected, a dialog box similar to that shown in Figure 4-6 is
presented to enter the desired value.
b)
Analog Output 4 Setup
- The radio button corresponding to the desired parameter to be
output by Analog Out 4 should be selected.
c)
SVAP Control -
The amount of sulfur vapor (S
vap
) detected in the tail gas indicates if the
sample probe’s cooling finger is at the optimum temperature. The probe’s cooling finger
temperature is controlled by using a solenoid valve to direct the flow of instrument air
through the finger off and on. The cooling finger temperature can be optimized in 2 ways:
•
By Svap – The concentration of sulfur vapor is monitored. If the concentration goes
above a pre set limit, then instrument air will be introduced to the probe’s cooling finger.
•
By RTD – The temperature of the tail gas exiting the probe is monitored. If the
temperature goes above a pre set limit then instrument air will be introduced to the
probe’s cooling finger.
4.8.2 Calculation Sub-Panel
The
Calculation
sub-panel (Figure 4-10) is used to set a broad range of calibration
parameters.
Figure 4-10: Calculation Sub-Panel
Содержание 943-TGX
Страница 48: ...Revision 4 48 August 02 2019 Figure 4 13 Network Sub Panel...
Страница 94: ...Revision 4 94 August 02 2019 Figure 9 3 Control Cabinet Detail...
Страница 96: ...Revision 4 96 August 02 2019 Figure 9 5 AC Wiring Diagram...
Страница 97: ...Revision 4 97 August 02 2019 Figure 9 6 DC Signal and Wiring Diagram...
Страница 98: ...Revision 4 98 August 02 2019 This Page Intentionally Blank...