Revision 5
23
August 2021
The steam jacketed ball valve must be installed on the process sample point
before mounting the oven cabinet. The Steam jacketed ball valve must remain
closed until the sample probe is installed and connected in the sample
handling oven
The oven cabinet should be mounted in a location where it is convenient to access to
make adjustments to the flow control valves as necessary and to lift off the oven cabinet
lid as required for maintenance procedures.
2.4.1.2 Mounting of the Control Cabinet
The framework of the control cabinet must be securely attached to a support structure at
the mounting site. The control cabinet must be mounted a distance away from the oven
cabinet such that the length of path for the fibre optic cables between the control cabinet
and the oven cabinet does not exceed 6 metres. The control cabinet should be mounted
in a location such that the display screen on the cabinet door is not mounted too high or
too low to be comfortably viewed by an operator, and also such that the cabinet door can
be opened to at least 90 degrees open. See Figure 3 for mounting details.
Figure 3: 943-TGS-CE Control Cabinet Dimensions
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...