5.6.
Battery autonomy estimations
Measuring channels
Radio settings
Firmware
version
Approx.
Autonomy *
3D
Accelerometer
3D Gyroscope
Barometric
pressure
Master / slave /
no radio
Version (release
date)
Hours
256Hz
256Hz
64Hz
-
1.0.2
(02.03.2017)
18h
128Hz
128Hz
64Hz
Master / slave
pairs / no radio
1.1.0
(14.06.2017)
20h
128Hz
128Hz
64Hz
Slave without
master
1.1.0
(14.06.2017)
10h
256Hz
256Hz
64Hz
Master / slave
pairs / no radio
1.1.0
(14.06.2017)
18h
256Hz
256Hz
64Hz
Slave without
master
1.1.0
(14.06.2017)
10h
512Hz
512Hz
64Hz
Master / slave
pairs / no radio
1.1.0
(14.06.2017)
15h
512Hz
512Hz
64Hz
Slave without
master
1.1.0
(14.06.2017)
10h
*Autonomy measured at room temperature
5.7.
Troubleshoot
Soft reset
The soft reset is the manipulation which can be done if the Physilog® is not working properly. To do
a soft reset, hold the main button for 15 seconds. The LED shines green, turns off and finally blinks
red/orange and then stops. If the problem persists, fill out the contact form on the support page
(www.gaitup.com/support) or write an email to: [email protected]. Please indicate Physilog®
serial number and describe the problem in as much detail as possible (number of red blinks, last
manipulation, firmware version etc.).
Hard reset
If the soft reset doesn’t allow to recover Physilog®, a hard reset is necessary.
A hard reset is done
when
the sensor’s battery
goes completely out (few days to several weeks depending on firmware
version and charging state). To force a hard reset, introduce a thin item (like the end of a paperclip)
into the hole besides the USB entry and gently press until you feel a button pressed inside the sensor.
W
arning:
This manipulation destroys the waterproofing membrane and the sensor is therefore no