8.4. Guidance and manufacturer’s declaration - Electromagnetic emissions
The Gait Up GO medical device has been developed for electromagnetic compatibility
according to IEC 60601-1-2 international standard.
The Gait Up GO medical device is intended for use in the electromagnetic environment
specified below. The user and/or end-user of the Gait Up GO medical device should ensure it is
used in such environment
Emissions test
Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
Gait Up GO medical device is a medical
electrical equipment which uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
Gait Up GO medical device is intended to be
used in home environment
Harmonic emissions
IEC 61000-3-2
N/A
N/A
Voltage fluctuations/
Flockers emissions
IEC 61000-3-3
N/A
N/A
Table 9. Electromagnetic emissions
8.5. Guidance and manufacturer’s declaration - Electromagnetic immunity
Immunity test
IEC 60601 test level Compliance level
Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 15 kV Air
+/- 8kV Contact
+/- 15 kV Air
+/- 8kV Contact
Floors should be
wood, concrete
or ceramic tile. If
floors
are covered with
synthetic
material, the relative
humidity
ENGLISH
46/51