6. T
ROUBLESHOOTING
What do you
see/hear
Problem
What to do
More Information
The Gait Up GO
mobile
application
displays an error
pop-up after you
select a test,
indicating that it
cannot connect
to the Physilog6
sensor device
There is a
problem with
the
connection to
the Physilog6
sensor device
Make sure that the
sensor device
accessory is in
proximity of your
mobile device.
Control that the sensor
device accessory has
enough battery to be
connected to the
application.
Click on the sensor
information button (
) to check that the
linked sensor device
accessory corresponds
to the sensor in your
possession (last four
digits of serial number
on the back of the
sensor corresponding
to the number
indicated in the mobile
application).
If the linked sensor device
does not correspond to
the sensor in your
possession, contact your
healthcare professional
who has prescribed the
device.
The Gait Up GO
mobile
application
displays an error
pop-up after you
select a test,
indicating that
the sensor
device accessory
has
incompatible
firmware
The Physilog6
sensor device
accessory has
incompatible
firmware
version.
Update the sensor
device firmware by
pressing the "Update
Physilog" button.
This error can appear
during the first use after
you update the Gait Up GO
mobile device or after
installation of the
application on a new
mobile device.
ENGLISH
37/51