MIC TRAC MT-3000
10
2010 Gagemaker, LP
RCDMMT300010-10
11. Adjust the height of the depth gage support to
hold the end of the depth rod, so it does not
bend during calibration. Tighten the thumb
screw.
12. Turn the coarse adjust knob counterclockwise
until the depth micrometer rod barely touches
the flat face anvil. Secure the coarse adjust
lock.
13. Continue turning the fine adjust knob until the
. in the axis display lights up. Record any
deviations on the Gage Calibration Record or
in-house calibration report. Continue with the
same calibration process for all remaining
depth rods.
14. Remove the depth micrometer and continue
with the same calibration process for the next
depth micrometer.
15. After calibrating all depth micrometers, be sure
to remove the depth gage fixture and flat face
anvil from the MT-3000. Oil the fixture and anvil
and return them to the storage case.
Calibration