CompuScope 1610C
147
Windowed triggering
Using the CompuScope 1610C’s ability to have two independent trigger sources which are ORed together, the
user can set up the triggering such that a trigger will occur if the input signal is outside a specified “window,” i.e.
it is higher than the upper limit or lower than the lower limit.
Figure 86: Windowed triggering
Channel 1 or 2 triggering
There are occasions when it is not possible to predict if the trigger is going to occur on Channel 1 or 2. For
example, if the CompuScope 1610C is hooked up to two photon detectors that are picking up random or
intermittent pulses, the user is not able to select the trigger source.
The CompuScope 1610C can be put in a mode in which it will trigger on, say, the rising edge of a pulse on
Channel 1 or on the rising edge of a pulse on Channel 2.
In other words, by selecting two different sources with independent levels and slopes, the user can generate a
trigger on the first occurrence of the pulse on either of the two channels.
Trigger bus for Master/Slave systems
In a Master/Slave system, any one of the input channels can be used as a trigger source.
This is achieved by having a trigger bus in the Master/Slave Timing Module that causes any one of the
CompuScope 1610C cards in the system to cause the entire system to trigger.
This powerful feature allows the use of a multi-channel, Master/Slave CompuScope 1610C system for acquiring
signals in a system in which the user cannot predict which channel would receive the trigger signal. Examples of
these applications are explosion testing and acoustic emission testing.
External Trigger amplifier
The External Trigger input also has oscilloscope-like 1 M
Ω
/ 30 pF input impedance.
Gage’s proprietary input-protection circuitry allows the FET stage to withstand voltages as high as
±
15 Volts
without causing any damage to the inputs.
The trigger comparators have an operating range of
±
1 Volt. Software-selectable attenuation of X1 or X5 on the
External Trigger input allows input ranges of
±
1 Volt and
±
5 Volt, respectively.
Содержание CompuScope 14100C
Страница 6: ......
Страница 24: ...18 Installing CompactPCI PXI Master Slave multi card systems Notes ...
Страница 30: ...24 Memory organization on CompuScopes except for CS85GC Notes ...
Страница 75: ...Driver installation guide 69 Help About CsTest ...
Страница 76: ...70 Driver installation guide Notes ...
Страница 85: ...CompuScope 85GC 79 CompuScope 85GC simplified block diagram Figure 52 CS85G simplified block diagram ...
Страница 90: ...84 CompuScope 85GC Notes ...
Страница 101: ...CompuScope 82GC 95 CompuScope 82GC simplified block diagram Figure 57 CS82GC simplified block diagram ...
Страница 124: ...118 CompuScope 14100C CompuScope 14100C simplified block diagram Figure 68 CompuScope 14100C simplified block diagram ...
Страница 146: ...140 CompuScope 1610C CompuScope 1610C simplified block diagram Figure 79 CompuScope 1610C simplified block diagram ...
Страница 165: ...CompuScope 3200C 159 CompuScope 3200C simplified block diagram Figure 89 CompuScope 3200C simplified block diagram ...
Страница 174: ...168 Notes Notes ...