Installation 

 

4.3

 

F

ITTING THE 

C

HART

 

 

  

Figure 13 - Chart Fitting 

 

To replace the chart, follow the steps:

  

 

 

Open the door of the recorder.

 

 

Unscrew the chart knob as shown in figure.

 

 

Remove the chart.

 

 

Insert the new chart.

 

 

Screw the knob after setting time axis. Make sure that chart slides below the clamp as shown 
in figure. 

4.4

 

R

EPLACE 

F

USE 

(B

ATTERY 

B

ACKUP

)

 

Follow the steps given below to replace the fuses:

  

 

 

Disconnect the recorder from the mains. This is very important to avoid shock hazard. 

 

Open the door of the recorder. 

 

Open the Chart plate. 

 

Use screwdriver to lever out the fuse holder. 

 

Replace the fuse with a new one of the proper rating (20mm, glass fuse, fast blow, 0.5A) 

 

Press the holder firmly back into place.

 

 

 

Содержание CR-2010 PRO Series

Страница 1: ......

Страница 2: ...10 4 1 2 Wiring diagram for recorder 12 4 2 Fitting the Pen 21 4 3 Fitting the Chart 22 4 4 Replace Fuse Battery Backup 22 Operation 23 Configuration 24 6 1 Configurable Item 24 6 2 Functions 26 6 2...

Страница 3: ...Pen 21 Figure 13 Chart Fitting 22 Figure 14 Two Pen Recorder Front Panel 23 Figure 15 Main menu display 25 Figure 16 Chart speed 26 Figure 17 Hour Minute option 26 Figure 18 Display Setting 27 Figure...

Страница 4: ...able 1 Manual Layout 4 Table 2 Sensor wiring 19 Table 3 Configurable Items 24 Table 4 Troubleshooting Guide 40 Table 5 Specification of Chart Recorder 42 Table 6 Order Code of Chart Recorder 46 Table...

Страница 5: ...ation This section describes the mechanical calibration procedure for the recorder Section 7 Troubleshooting Guide This section describes about the most frequently asked questions and their solutions...

Страница 6: ...w about installed options in your chart recorder 3 4 UNPACKING AND INSPECTION OF RECORDER G Tek recorders are dispatched in a recyclable environment friendly package specially designed to give adequat...

Страница 7: ...6 Introduction Figure 1 Unpacking and Inspection of Recorder...

Страница 8: ...7 Installation INSTALLATION Figure 2 Environmental Conditions and Overall Dimensions...

Страница 9: ...8 Installation Figure 3 Pipe Mounting Figure 4 Wall Mounting...

Страница 10: ...9 Installation Figure 5 Panel Mounting...

Страница 11: ...14AWG 2 5mm1 The instrument and all inputs and outputs conform to Mains Power Input Insulation Category ll All connections to secondary circuits must have basic insulation After installation there mu...

Страница 12: ...ion devices across the coil Coil transients can also be transmitted through the air so the recorder itself should be mounted as far as possible from power control devices and or wiring 4 When line pow...

Страница 13: ...eutral and Earth Should not be more than 6V If by any chance it is more than 30V it is a serious fault on earth line and must immediately be corrected Proper Earthing is essential for safety of the pe...

Страница 14: ...13 Installation Figure 6 Opened Chart Plate and Front View of Base Box Figure A Open Chart Plate Sensor PCB Figure B Front view of base box with power supply PCB mounted on it...

Страница 15: ...14 Installation 4 1 2 1 Mains supply connection Battery Terminal connection Relay connection Figure A...

Страница 16: ...le to the Earthing terminal given on body of the recorder as shown in figure B The connection of 12V DC battery is shown in figure 7 Connect the and of 12V battery to the and terminal of the connector...

Страница 17: ...zes the effects of RFI and ESD 3 Exceptional care should be taken when wiring to relay or solenoid coils as large transients are produced when coils or any other inductive devices are switched This pr...

Страница 18: ...iring The Recorder can be connected and configured to operate with a variety of signal sources as thermocouple RTDs dc current dc voltage etc The sensor wiring for different sensors for 2 Pen recorder...

Страница 19: ...types Figure 9 Motherboard PCB The Recorder can be connected and configured to operate with a variety of signal sources as thermocouple RTDs dc current dc voltage etc The sensor wiring for different s...

Страница 20: ...gth are equal and that each lead exhibits no more than 10ohm resistance between the recorder and the RTD Pt 100 3 For Input connections high quality low resistance contacts must be used which are suit...

Страница 21: ...on of chart recorder with Transmitter and External Power Supply is as Shown in figure 10 Connecting Recorder with transmitter using Internal Power Supply Figure 11 Connecting Recorder with transmitter...

Страница 22: ...ep 2 A new pen may have burrs or obstructions in the guides Apply sufficient force to clear the guides or use sharp knife to clean the guide beforehand Caution Improper fitting of the pen may result i...

Страница 23: ...Make sure that chart slides below the clamp as shown in figure 4 4 REPLACE FUSE BATTERY BACKUP Follow the steps given below to replace the fuses Disconnect the recorder from the mains This is very im...

Страница 24: ...ramming menu Refer Programming section Figure 14 Two Pen Recorder Front Panel KEYS There are five multifunction keys are available on the front panel of the chart recorder to configure the different p...

Страница 25: ...4 5000 0 to 5000 0mV 0 0 to 5000 0mV Channels to scan The parameter value for all channel digital inputs and digital outputs will be display with rolling action Also scan channel with Freeze action Ra...

Страница 26: ...Chart Full Scale and Chart Zero Communication Parameters Device Address Address of chart recorder for communication with computer Saving Configuration Channel parameters The chart recorder can be Conf...

Страница 27: ...hen Mn In St option seen in chart speed menu OR 6 2 1 1 Hour in step Minute in step If Chart speed is set in hour in step mode one hour is counted for single count Count range is 1 to 255 If Chart spe...

Страница 28: ...es not affect on pen movements 6 2 2 1 Freeze panel Use to select the panel and as on selected display action it will be display continuously or in rolling manner To set freeze panel follow the proced...

Страница 29: ...t offset for temperature reading a mathematical value is directly added to the reading coming on led display with the reference value which we consider as an accurate value coming on other display dev...

Страница 30: ...ark Pen 1 In pen park if we set pen 1 pen 1 will be shifted to full scale position Channel 1 on led display continue showing channel 1 current reading Other pens remain on their current reading positi...

Страница 31: ...d its position all operation will be stopped Display will be blink with restart After changing the pen user should be restarting the device 6 2 6 3 Change Pen 2 Same as Describe in Type of Functions 6...

Страница 32: ...31 Configuration Sensor Range Lo Chnl 1 Span Offset Multiply Resolutn Alrm SpH Alrm SPL 1 128 55 ChartZero ChartSpan Figure 25 Channel Configuration...

Страница 33: ...between resistance and temperature is relatively linear as shown below for a sample 100 RTD Figure 26 RTD Sensor 6 2 7 1 2 Thermocouple Recorder accepts as listed under TC J It is made up of iron ve e...

Страница 34: ...68 uV C TC N It is made up of nicrosil ve electrode nisil ve electrode It operates over the temperature range of 200 to 1300 C temperature Its sensitivity is 39 uV C 6 2 7 1 3 Process Inputs There is...

Страница 35: ...ure by pressing the appropriate buttons given on the top panel of the chart recorder Figure 27 Sensor Selection TC T TC B Save Save TC E TC N Save Save TC R TC S Save Save TC K TC C Save Save RTD TC J...

Страница 36: ...he chart recorder 6 2 7 3 Span Span is the total range of values that a user wants to measure Span value indicates the total interval of which all the output values will be displayed For example If th...

Страница 37: ...e output display For example If the output is 2 301 then according to user he can set the display output If Resolution is kept 0 then output is 2 If Resolution is kept 0 1 then output is 2 3 If Resolu...

Страница 38: ...nd low range respectively then internal relay will activate the Industrial buzzer make sound till value will not come back within range 6 2 7 9 Chart Zero This feature is used for selecting the minimu...

Страница 39: ...Total CCR will plot the outputs between the range of 0 to 100 Now for the above case if the user wants to just measure and plot the temperature above 50 degree that is 50 100 degree then in this case...

Страница 40: ...keys are used to set the pen up or down All the pens of their individual different channels can be given different mechanical calibration according to the desired output At Zero Tst pen moves on zero...

Страница 41: ...the channel to which the sensor input is provided Pen is not marking Pen is dry replace the Pen Check pen arm pressure Chart does not move Chart knob is loose Check chart spd Chart motor is having pro...

Страница 42: ...ESSORIES 9 1 STANDARD ACCESSORIES Charts Pack of 30 Pens Pack of 5 1 number 2 number for two channel Panel Mounting Clamps 2 numbers 9 2 OPTIONAL ACCESSORIES Wall Mounting Kit Case to Panel Gasket Fig...

Страница 43: ...Ranges Standard Refer Table 2 Customized Please specify Display Operator Panels and Input Display Type Digital 8 digits ultra bright LED Numeric semi alphabetic Display Height 0 3 Status Indicator Cha...

Страница 44: ...ted Environmental Temperature Operation 5 C to 45 C Limiting 0 C to 50 C Storage 20 C to 60 C Humidity Operation 10 to 80 RH Non Condensing Storage 5 to 90 RH Non Condensing Altitude 2000 meter Power...

Страница 45: ...05 Operations Load 1A 230V AC resistive Relay Refresh Rate 1 Sec Isolation Relay Relay 2 5KV 1Minute Isolation Relay Contact GND 1 5KV 1Minute Safety EMI EMC IEC 61326 1 Class A Safety IEC 61010 1 Pol...

Страница 46: ...00 K IEC584 1 200 To 1300 C Hoskins 0 To 2300 R IEC584 1 0 To 1750 S IEC584 1 100 To 1750 T IEC584 1 100 To 400 B IEC584 1 200 To 1800 E IEC584 1 100 To 1000 N IEC584 1 250 To 1300 mV Bipolar 1V 99999...

Страница 47: ...dth Recorder type Pen Display Power Supply Relay PC Interface Chart Type Table 7 Range Chart Speed Sensor Type CR PD PS RE PI CT R CS S 3 CR11 NU 1 1P D 0 85 265 V CE 0 None 0 None x x 0 PG 0 PG 0 Uni...

Страница 48: ...1 to 5 24H 11 301011 D110150 1 5 Dual Range 58 0 to 250 0 to 100 24H 11 301015 D110250 0100 Dual Range 59 0 to 100 0 to 400 24H 11 301004 D110100 0400 Dual Range 60 0 to 100 40 to 60 24H 11 301005 D11...

Отзывы: