Actuator Calibration Stand Type AL0010 - Page  9

G.R.A.S.  Sound & Vibration

Fig. 6

 

Block diagram of a complete set-up for making tests

Response signal  from test micro- phone to measuring  ampli

er

From signal generator to 

AC Input

Actuator Supply  Type 14AA

To 

AL0010 from 

AC Output  + 

800 

DC  (via 

thin lead of  AL0010)

AL0010 set up  for testing a  1⁄2-inch micro- phone (see  Fig. 

3)

Power Module  Type 12AK

Computer with hardware and  software for simulating:

Ÿ

 a signal generator capable  of generating logarithmi- cally swept tones

Ÿ

 a measuring ampli

er

From preampli

er to Input

Содержание AL0010

Страница 1: ...G R A S Sound Vibration Skovlytoften 33 2840 Holte Denmark Tel 45 45 66 40 46 Fax 45 45 66 40 47 Instruction Manual Actuator Calibration Stand Type AL0010 ...

Страница 2: ...Revision 02 11 2004 Actuator Calibration Stand Type AL0010 ...

Страница 3: ...Actuator Calibration Stand Type AL0010 Page 2 G R A S Sound Vibration Actuator Calibration Stand Type AL0010 Actuator Calibration Stand Type AL0010 ...

Страница 4: ... mounting a 1 inch microphone on a 1 2 inch preampli er Additional Equipment Fig 5 shows what additional equipment can be supplied by G R A S Sound Vibration and Fig 6 shows a block diagram of a possible set up with all additional equipment The following is a list of all required additional equipment 1 Power supply for the 1 2 inch preampli er e g the G R A S Power Module Type 12AK shown in Fig 5 ...

Страница 5: ... swept tone Items 3 and 4 could be combined in the same unit e g a computer tted with suitable hardware and software for A D and D A conversions in order to simulate both a signal generator and measurement ampli er Test Procedure General The basic stages in the test procedure are 1 Setting up the microphone on the AL0010 see Fig 3 or Fig 4 2 Applying a signal via the actuator supply to the electro...

Страница 6: ...l output level as required Applying a Swept Signal and Measuring the Output from the Microphone The following describes a typical procedure for applying the swept signal It is assumed that the generator and measuring ampli er work to produce constant con dence results i e maintaining a constant β T product in real time throughout the frequency range of interest and make the measurement data availa...

Страница 7: ... the frequency response of a 1 2 inch condenser microphone 1 2 inch Electro static Actuator RA0014 1 2 inch condenser microphone without protection grid Remove protec tion grid Parking places for 1 2 inch Actuator when not in use 1 2 inch preampli er From the 1 2 inch Electro static Actuator RA0014 ...

Страница 8: ...he frequency response of a 1 inch condenser microphone 1 inch Electro static Actuator RA0015 1 inch condenser microphone without protection grid Remove protec tion grid Adapter RA0017 Parking places for 1 inch Actuator when not in use 1 2 inch preampli er From the 1 inch Electro static Actuator RA0015 ...

Страница 9: ...Actuator Calibration Stand Type AL0010 Page 8 G R A S Sound Vibration Fig 5 Showing AL0010 in a set up with Power Module Type 12AK and Actuator Supply Type 14AA Type 14AA Type 12AK ...

Страница 10: ...g ampli er From signal generator to AC Input Actuator Supply Type 14AA To AL0010 from AC Output 800 V DC via thin lead of AL0010 AL0010 set up for testing a 1 2 inch micro phone see Fig 3 Power Module Type 12AK Computer with hardware and software for simulating a signal generator capable of generating logarithmi cally swept tones a measuring ampli er From preampli er to Input ...