1
"
Pressure
Microphone
Type 40EN
1
"
to ½
"
Adapter
RA0017
½
"
to ¼
"
Adapter
RA0003,
NOT Included
with the
¼
"
Preamplifier
Type 26AS
¼
"
Preamplifier, Short
Type 26AS *
1
"
to ½
"
Microphone
Adapter
RA0058
1
"
Microphone
Protection
Grid
RA0177
½
"
to ¼
"
Adapter
GR0010
(included with the
¼
"
Preamplifier)
½
"
Pressure
Microphone
Type 40AP,
protection
grid removed
½
"
Pressure
Microphone
Type 40AD,
protection
grid removed
¼
"
Preamplifier, Short
Type 26CS *
Cable AA0018
G.R.A.S. SOUND & VIBRATION
Page 8
Hearing-Protector Test Fixture Type 45CA
Fig. 4.1 Assembly options (exploded view)
4
Configuration with Pressure Microphone,
1
"
Type 40EN or ½
"
Type 40AP/40AD
Applications
Insertion Loss of Ear Muffs. If Microphone 1
"
Type 40EN is used, the configuration is in conform
-
ity with ISO 4869-3.
Required Accessories
•
The parts shown in Fig. 4.1 (choose one of the three assembly options)
•
Pistonphone Type 42AP (recommended) or Type 42AA
•
Test instruments such as those shown in Fig. 4.14
4.1
Assembling Microphone and Preamplifier
Fig. 4.1 displays three options for the microphone/preamplifier assembly required for the meas
-
urements.
1
"
Type 40EN OR ½
"
Type 40AP OR ½
"
Type 40AD
_____________
* For space limitations, the super-short ¼
"
Preamplifier Type 26AS/Type 26CS (shown) is required for binaural measure-
ments. For monaural measurements, replaceType 26AS with Type 26AC and Type 26CS with Type 26CB.
Содержание 45CA
Страница 2: ...Revision 28 November 2013 LI0024 ...