13
LI0083 – Revision 22 June 2017
OVLLED Y<cr>
set the conditioning amplifier to display overload.
(default).
OVLLED N<cr>
set the conditioning amplifier to not display overload.
(not recommended).
Speaker Amplifier
SG-20<cr>
set the speaker amplifier gain to –20 dB.
SG-10<cr>
set the speaker amplifier gain to –10 dB.
SG0<cr>
set the speaker amplifier gain to 0 dB.
SG10<cr>
set the speaker amplifier gain to 10 dB.
SMUTE Y<cr>
mute speaker amplifier.
SMUTE N<cr>
enables speaker amplifier.
SHP0<cr>
DC couples speaker amplifier.
SHP10<cr>
AC couples speaker amplifier HP 10 Hz.
Overload Latching
SLATCH<cr>
set the speaker amplifier to latch overloads
overload’s are latched, and cleared when read.
No overload messages will be sent.
Gain
CG0<cr>
set the speaker amplifier gain to 0 dB.
CG20<cr>
set the speaker amplifier gain to 20 dB.
Current Overload Latching
CLATCH<cr>
set the speaker amplifier to latch current overloads.
Overload’s are latched, and cleared when read. No
overload messages will be sent.
Fast Settling
FastSettling #<cr>
Set the period of fast settling to [#] sec. Enabling
fast settling after change of setup in # sec. (Inserts
HP50Hz for [#] sec.)
FastSettle<cr>
Enables fast settling in the predefined period, the period
is defined with the command
FastSettling #