- 20 -
APPENDIX II
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
rms
a
0.15 MHz - 80 MHz
6 V
rmsa
in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
b
80% AM at 1 kHz
3 V
rms
a
0.15 MHz - 80 MHz
6 V
rmsa
in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
b
80% AM at 1 kHz
Radiated RF EM
fields
IEC 61000-4-3
10 V/m
c
80 MHz - 2.7 GHz
80% AM at 1 kHz
10 V/m
c
80 MHz - 2.7 GHz
80% AM at 1 kHz
The
Sphygmomanometer
(MD4120/MD4140/MD4180)
is intended for use in the electromagnetic
environment specified below. The customer of the user of
Sphygmomanometer
(MD4120/MD4140/MD4180)
should assure that it is used in such an environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the
Sphygmomanometer (
MD4120/MD4140/MD4180
),
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the
transmitter in Watts (W), d is the minimum
recommended separation distance in meters (m), and
E is the immunity test level in V/m.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Immunity test
Compliance level
Electromagnetic environment - guidance
IEC 60601
test level