Step 2: Calibrate
of A to D converter.
Send a span signal to terminal 4 and 5 with correct polarity.
The span signal is 60 mV for thermocouple input, 1V for
0-1V input, 10V for 0-10V input and 20mA for 0-20 mA input.
Press
for at least 4 seconds. The display will blink a
moment. If the display didn't blink, then the calibration failed.
Gain
Step 3: Calibrate
of
.
Setup the equipment according to the following diagram
for calibrating the cold junction compensation. Note that a
K type thermocouple must be used.
offset
cold junction
offset
cold junction
4
5
Set the calibrator to be configured as K type thermocouple
output. Calibrator must have an internal compensation.
Send a 0.00 C signal to the unit under calibration.
Calibrator
K-TC
K+
K
NOTE:
The unit under calibration is powered in a still-air room
at a temperature
. Allow at least 20 minutes for
warming up.
25 +/-3 C
NOTE:
The unit under calibration is powered in a still-air room
at a temperature 25 +/-3 C. Allow at least 20 minutes for
warming up.
Figure 5-2
Cold Junction
Calibration Setup
Figure 5-2
Cold Junction
Calibration Setup
45
L 9 1
Stay at least 20 minutes in still-
air room room temperature
25 +/- 3 C
UM L91-Rev 8
With CJTL on the display adjust the value to 0.00 reading.
Once adjusted, Press
for at least 4 seconds. The display will
blink a moment. If the display didn't blink, then the calibration
failed.
The L91 being calibrated for Cold Junction Compensation MUST
be programmed for K t/c input, Celsius display
to performing
the CJTL calibration.
prior