When handling the sensor, do not pinch the ends 
where the load is applied.

When installing the sensor, connect it to an instrument 
and monitor the output to prevent possible overload. 

Do not apply excessive torque or create a torque 
through the sensor when attaching the fixtures.

In an environment with a high amount of moisture or 
humidity, create a drip loop on the cable to prevent 
any water from flowing into the sensor.

For compression, install the sensor right side up for 
best results (reference the label).  For tension, install 
the sensor upside down so the load is being applied 
to the measuring end.

When routing the cable to an instrument, you want to 
create a service loop in the cable so it is not taut or 
stretched. (Cable bend radius of 0.80" should be 
maintain where possible.) 