Insertion Loss Test Kit

Содержание OFS P10188-01

Страница 1: ...For Use With 50 62 5 200 400 m HCS Cable Assemblies ST SMA V Pin and F07 OFS Connectors Your Optical Fiber Solutions Partner Insertion Loss Test Kit...

Страница 2: ...dirty connector ends Always replace the protective dust caps onto the optical ports when not in use WARRANTY One year Limited Warranty OFS Test Kits are warranted against defective material and workm...

Страница 3: ...essories vi vii Function Definitions and Descriptions 1 Insertion Loss Testing Overview Single Ended Method 4 5 Substitution Method 6 7 Content Page Insertion Loss Test Procedure Step 1 8 Step 2 9 Ste...

Страница 4: ...nder Test not included Launch Jumper V Pin Connectors Pictured Parts Vary by Test Set Connector Type and Size All kits include Light Source and Power Meter Quick Start Guide and Owner s Manual Each ki...

Страница 5: ...or this HCS fiber size with these connectors Part Number 50 m ST P10188 10 62 5 m ST P10188 10 200 m V Pin P10188 01 400 m V Pin P10188 02 200 m ST P10188 03 400 m ST P10188 04 200 m SMA P10188 05 400...

Страница 6: ...y 25 C and 10 dB m 0 25 dB Resolution 0 01 dB General Specifications Operating temperature range 0 C to 50 C Storage temperature range 30 C to 60 C Relative humidity range 0 to 95 non condensing Dimen...

Страница 7: ...scription P10199 01 V Pin Adapter Cap P10199 02 ST Adapter Cap P10199 03 SMA Adapter Cap P10199 04 F07 Adapter Cap AP04707 V Pin Splice Bushing AP01960 ST Splice Bushing AP01618 SMA Splice Bushing AP0...

Страница 8: ...Set P10198 10 ST 50 m Launch Jumper P10198 08 ST 62 5 m Launch Jumper P10198 03 ST 200 m Launch Jumper P10198 04 ST 400 m Launch Jumper P10198 05 SMA 200 m Launch Jumper P10198 06 SMA 400 m Launch Ju...

Страница 9: ...m kit only Gray 50 m Black 62 5 Splice Bushing Launch Jumper Adapter Cap Light Source Device Under Test not included 1 Function Definitions and Descriptions V Pin Connectors Pictured Parts Vary by Te...

Страница 10: ...launch jumper must be the same as the fiber type of the device under test See Related Products and Accesso ries s ection for available launch jumper types Adapter Cap Screws onto detector head of Pow...

Страница 11: ...witches the measurement mode from absolute power dBm to loss dB OFS procedures outlined in this manual for simple insertion loss measurements utilize the dB mode Ref Button Pressing and holding down t...

Страница 12: ...s are configured for Substitu tion Method All other kits are configured to use the Single Ended Method For 50 and 62 5 m testing only an ad ditional step is required Wind the launch jumper around the...

Страница 13: ...end X of the device under test via a splice bushing Connector end Y of the device under test is connected to the detector See Figure 2 The detector displays dB loss which represents optical power loss...

Страница 14: ...The detector is zeroed to eliminate the effects of loss through the launch jumper and source Figure 3 NOTE V System kits are configured for Substitu tion Method All other kits are configured to use th...

Страница 15: ...ctor ends X and Y of the device under test are connected to Splice Bushings A and B respectively see Figure 4 The detector displays dB loss which represents optical power loss at the X connection due...

Страница 16: ...up on the toggle switch and moving to left or right to select the appropriate wavelength NOTE To turn OFF light source pull up on tog gle switch and move to center position Toggle Switch and Wavelengt...

Страница 17: ...yed Remove protective caps from Source and Meter Install appropriate Adapter Cap onto the Power Meter Allow 2 minutes for the test set to stabilize 9 NOTE To disable the 5 minute auto off feature hold...

Страница 18: ...Meter Connect the two Launch Jumpers with a Splice Bushing Zero the Power Meter by pressing and holding down the Ref button until HELD and 0 00 dB is displayed NOTE It is recommended to repeat this ze...

Страница 19: ...n the Power Meter display represents the insertion loss due to the connector at the Splice Bushing plus the attenuation through the length of cable For Substitution Method See Figure 4 on page 7 Disco...

Страница 20: ...s damaged re terminate or replace Replace 9 volt battery Low battery on light source or power meter Bat indicates need for battery replace ment Unable to plug in device under test to test unit or spli...

Страница 21: ...r Keep all optical ends clean to avoid poor insertion loss read ings and also to prevent contamination of the optical ports and ends Always replace the protective dust caps onto the optical ports when...