Funke Gerber SuperVario - N Скачать руководство пользователя страница 3

User Manual SuperVarioN

Page 5 of 15

   

              Issue dd. 28.07.2004

The temperature can be adjusted in steps of 0,5 °C and the centrifuging
time in minutes. All settings are non-volatile. They remain stored even
in case the centrifuge is switched off or the mains cable is disconnected.
All settings can be made only when centrifuge is in standby  (not while
centrifuging).

2.2.2.  Setting to operation

After having assembled and prepared the centrifuge according to point
1 of this manual the display is showing the desired speed (revolutions
per minute), the actual temperature (in oC) and the desired centrifuging
time by the active mode. Take care that the device is set up according
to the desired analysis method, i.e. the values  (as per point 3) are set
properly.

2.2.3.  Start of centrifuge

Attention: Recheck the right fastening of the tommy screw !
The test bottles have to be placed symmetrically to avoid unbalances.
Afterwards the lid is closed and pressed down until an audible catching
can be heard. Then press the key START and start centrifuging.
In case the lid was erroreously closed and the centrifuge has not yet
started it can be opened by pressing the key STOP/OPEN.
At the end of centrifuging time the centrifuge is braking and opening
the lid locking. The braking time depends on the speed. It is between
about 15 sec. (at 600 rpm) and 25 sec. (at 1100 rpm) Braking times are
preset.
The centrifuge switches off automatically in case it is out of balance and
the free moveable motor is touching the housing.
Hint: On account of safety reasons every switching off due to unbalance
is followed by a braking. Same is done in case of a coarse knocking on

User Manual SuperVarioN

Page 6 of 15

   

              Issue dd. 28.07.2004

the centrifuge. In order to stop the centrifuge by hand press the key
START a second time. 

Hint:

 The centrifuge can only be started with lid locked and in case it is

not in the setting modus.

Attention:

 The centrifuge bowl is to be kept always clean. In case

glasses were broken it should be cleaned immediately

3. Settings

3.1.   Setting of Speed

First press key "RpM/g". Then use the keys marked with arrow to find
the new desired speed.
The desired speed is variable between 600 and 1130 rpm in steps of 10.
Press the key "RpM/g" a second time in order to store the desired value.

3.2.  Setting of Temperature

First press key "°C". The display is showing the set temperature. It can
be altered by the two keys marked with arrow.
By pressing the key "°C" a second time the desired temperature is
stored and the display is showing again the actual temperature.
Notice! During standby the centrifuge reduces the heating to avoid local
overheat. The amount of reduction is 10 °C.

3.3.  Setting of Centrifuging time

First press key "TIME". By using the two keys marked with arrow the
new desired time can be set.
By pressing the key "TIME" a second time this setting is stored.

Содержание SuperVario - N

Страница 1: ...g 3 2 Using the centrifuge 3 2 1 Function of keys 3 2 2 Operation 4 2 2 1 Display Button board 4 2 2 2 Setting to operation 5 2 2 3 Start of centrifuge 5 3 Settings 6 3 1 Setting of speed 6 3 2 Settin...

Страница 2: ...otecting the motor during transportation Possible residues of adhesive tape have to be removed Put the desired head on the motor spindle and fasten it by the tommy screw and the washer The tommy screw...

Страница 3: ...rpm and 25 sec at 1100 rpm Braking times are preset The centrifuge switches off automatically in case it is out of balance and the free moveable motor is touching the housing Hint On account of safet...

Страница 4: ...be done only in case of emergency The cap is mostly destroyed after it and has to be exchanged Using the centrifuge User Manual SuperVarioN Page 8 of 15 Issue dd 28 07 2004 with cord lying open is fo...

Страница 5: ...y analyses For Gerber method set the speed to 1100 RpM For method according to Roese Gottlieb Mojonnier tubes set the speed to 600 RpM For ADMI methode set the speed to 900 RpM That values are valid o...

Страница 6: ...87 Head C up to 6 solubility index tubes 3631 Bucket for butyrometer of aluminium alloy User Manual SuperVarioN Page 12 of 15 Issue dd 28 07 2004 3632 Bucket for Babcock bottles 3633 Bucket for solubi...

Страница 7: ...User Manual SuperVarioN Page 13 of 15 Issue dd 28 07 2004 User Manual SuperVarioN Page 14 of 15 Issue dd 28 07 2004...

Страница 8: ...safety and health We declare further that all Instruments are strictly tested by our internal quality test procedure Each Instrument has passed the end production control In case of modification in us...

Отзывы: