Funke Gerber SuperVario - N Скачать руководство пользователя страница 1

User Manual SuperVarioN

Page 1 of 15

   

              Issue dd. 28.07.2004

SuperVarioN

Multi Purpose  Centrifuge

SuperVario - N

Funke- Dr. N. Gerber Labortechnik GmbH

D - 12105 Berlin,  Ringstraße 42

Tel.: +49 - 30 - 702 006 - 0     Fax: +49 - 30 - 702 006 - 66

User Manual SuperVarioN

Page 2 of 15

   

              Issue dd. 28.07.2004

Contents

1  

Assembling and starting

3

2  

Using the centrifuge

3

2.1 

Function of keys

3

2.2 Operation

4

2.2.1

Display/Button board

4

2.2.2 

Setting to operation 

5

2.2.3 

Start of centrifuge 

5

3 Settings

6

3.1  

Setting of speed 

6

3.2 

Setting of temperature

6

3.3 

Setting of centrifuging time

6

4  

Informations about using

7

5

Tabular  summary on Relative 
Centrifugal Acceleration

10

6  

Technical data

11

7 Ordering 

data

11

8 Spare 

parts

12

Annex 1:  Wiring diagram main control

13

Annex 2:  Wiring diagram motor control

14

Annex 3:  Assembling plan

15

Содержание SuperVario - N

Страница 1: ...g 3 2 Using the centrifuge 3 2 1 Function of keys 3 2 2 Operation 4 2 2 1 Display Button board 4 2 2 2 Setting to operation 5 2 2 3 Start of centrifuge 5 3 Settings 6 3 1 Setting of speed 6 3 2 Settin...

Страница 2: ...otecting the motor during transportation Possible residues of adhesive tape have to be removed Put the desired head on the motor spindle and fasten it by the tommy screw and the washer The tommy screw...

Страница 3: ...rpm and 25 sec at 1100 rpm Braking times are preset The centrifuge switches off automatically in case it is out of balance and the free moveable motor is touching the housing Hint On account of safet...

Страница 4: ...be done only in case of emergency The cap is mostly destroyed after it and has to be exchanged Using the centrifuge User Manual SuperVarioN Page 8 of 15 Issue dd 28 07 2004 with cord lying open is fo...

Страница 5: ...y analyses For Gerber method set the speed to 1100 RpM For method according to Roese Gottlieb Mojonnier tubes set the speed to 600 RpM For ADMI methode set the speed to 900 RpM That values are valid o...

Страница 6: ...87 Head C up to 6 solubility index tubes 3631 Bucket for butyrometer of aluminium alloy User Manual SuperVarioN Page 12 of 15 Issue dd 28 07 2004 3632 Bucket for Babcock bottles 3633 Bucket for solubi...

Страница 7: ...User Manual SuperVarioN Page 13 of 15 Issue dd 28 07 2004 User Manual SuperVarioN Page 14 of 15 Issue dd 28 07 2004...

Страница 8: ...safety and health We declare further that all Instruments are strictly tested by our internal quality test procedure Each Instrument has passed the end production control In case of modification in us...