VPAD Manual
43/47
25. VPAD H standard spindle selection table
Maximum guideline values in Poise
RPM/SP
R1
R2
R3
R4
R5
R6
R7
0.3
2.6K
10.6K
26.6K
53.3K
106.6K
266.6K
1M
0.5
1.6K
6.4K
16K
32K
64K
160K
640K
0.6
1.3K
5.3K
13.3K
26.6K
53.3K
133.3K
533.3K
1
800
3.2K
8K
16K
32K
80K
320K
1.5
533.3
2133
5.3K
10.6K
21.3K
53.3K
213.3K
2
400
1.6K
4K
8K
16K
40K
160K
2.5
320
1.28K
3.2K
6.4K
12.8K
32K
128K
3
266.6
1066
2.6K
5.3K
10.6K
26.6K
106.6K
4
200
800
2K
4K
8K
20K
80K
5
160
640
1.6K
3.2K
6.4K
16K
64K
6
133.3
533.3
1.3K
2.6K
5.3K
13.3K
53.3K
10
80
320
800
1.6K
3.2K
8K
32K
12
66.6
266.6
666
1.3K
2.6K
6.6K
26.6K
20
40
160
400
800
1.6K
4K
16K
30
26.6
106.6
266
533
1066
2.6K
10.6K
50
16
64
160
320
640
1.6K
6.4K
60
13.3
53.3
133.3
266.6
533
1.3K
5.3K
100
8
32
80
160
320
800
3.2K
Table 20.
ATTENTION:
K Indicates thousands.
Example: 7.8K = 7800
M Indicates millions
Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of the
selected scale.