VPAD Manual
7/47
5.
Specifications
Speeds:
0.01 – 250 RPM
Temperature Sensing Range:
-40 °C to 300 °C (-40 °F to 572 °F)
USB A Port for use with temperature probe
Viscosity Accuracy:
±1.0 % of full scale range
The use of accessory items will have an effect on the measurement
accuracy
Viscosity Repeatability:
±0.2 % of full scale range
Temperature Accuracy:
±0.1 °C | -40 °C to 300 °C (-40 °F to 572 °F)
Operating Environment:
+5 °C to 40 °C temperature range (41 °F to 104 °F)
Non condensing humidity
Electrical Certifications:
Conforms to CE Standards:
EN 61010:
Safety requirements for electrical equipment, for measurement, control and
laboratory use
EN 61326-1:
Electrical equipment for measurement, control and laboratory use
Notice to customers:
This symbol indicates that this product is to be recycled at an appropriate collection
center
6.
Conditions for use
-
Indoor use
-
Maximum altitude 2000 m.
-
Surrounding temperature range: from +5 to 40ºC.
-
The equipment temperature must be kept above the dew point so moisture doesn't condense on or
in it. The power source fluctuations should not surpass
10% of the nominal voltage
-
Installation category II
-
Pollution level II
7. Maintenance
Always clean all of the parts after each use! Clean and dry the spindles and the spindle guard well.
Make sure that there is not any sample remaining, especially in the delicate zones such as the
spindle connector.
Use detergents or solvents to clean the spindles and the protector:
-
For cleaning food samples, use lukewarm water and if necessary, use soft household
detergents