14-17
Chapter 14 Troubleshooting
Troubleshooting
Recorder Unit (HR-800)
"Check extended memory: Turn
OFF the power, and turn it ON
again after 5 seconds." is
displayed.
The system cannot recognize the
internal memory.
Turn OFF the power of the DS-8900 main unit, and
then turn it ON again after 5 seconds. If the same error
persists, refer to your nearest service representative.
"Check extended memory" is
displayed.
The system cannot recognize the file
system of the internal memory.
Turn OFF the power of the DS-8900 main unit, and
then turn it ON again after 5 seconds. If the same error
persists, refer to your nearest service representative.
Situation
Cause
Solution
The "Check Paper" message is
displayed.
There is no paper.
Set the paper in the paper holder.
The "Check Cassette" message is
displayed.
The paper holder is open.
Close the cassette until it locks into place with a click
sound.
The "Check Paper" or "Check
Cassette" message is not
displayed, but printing cannot be
performed.
The paper is not correctly installed. The
front and backside of the paper is set
oppositely.
Set the paper in the paper holder so that the logo,
FUKUDA DENSHI CO.,LTD appears on the upper
surface.
(
"Installing the Recording Paper (Optional)" P5-3)
Only the ECG waveform is printed.
The waveforms to be printed is not set
under "Print Settings".
Set the printing waveforms for manual, alarm, periodic
printing for each bed.
(
"Printing Condition/Output Destination Setup" P12-
Alarm printing does not function.
The alarm printing mode is set to OFF. Set [ON] for "Alarm Printing". (Each Bed > Print Settings
> Alarm Printing) Also, set the printing waveforms and
alarm factors.
(
Alarm is set to OFF for the parameter.
On the alarm setup menu for the corresponding
parameter, set the alarm [ON]. Also, set the upper and
lower alarm limit. (
Parameter" P7-4)
Periodic printing does not function.
The periodic printing mode is set to
[OFF].
Set [Printer] or [Recall] for "Periodic Printing". (Each Bed
> Print Settings > Periodic Printing) Also, set the printing
waveforms and periodic interval/timer.
(
"Periodic Printing Setup" P12-6)
Telemetry remote printing does not
function.
The event button on the transmitter is
not pressed long enough to transmit
the signal to this equipment.
Press the event button for more than 3 seconds.
The telemetry remote printing function
is set to [OFF].
Select [ON] for "LX Remote Printing". (Initial Settings >
User I/F > Display/Print)
(
Maintenance Manual "Display/Print" P7-12)
Remote printing does not function.
A recorder is equipped on the bedside
monitor.
Perform the printing on the bedside monitor. If a recorder
is equipped on the bedside monitor, telemetry remote
printing will not function.
The "Check Printer" message is
displayed.
The thermal head temperature has
increased.
Damage to the thermal head can be considered. Contact
your nearest service representative.
Situation
Cause
Solution
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......