8-5
Chapter 8 Parameter Setup
ECG
below -20 mm / -2.0 mV is set, the lower alarm will turn OFF.
Size / Lead
Adjustment of Waveform Size and Baseline Position
Adjust the waveform size and baseline position.
CAUTION
Automatic size of the ECG is effective only at the time the [Auto] key is pressed. This does
not continually adjust the size.
The waveform size cannot be set if the waveform is not displayed. Change the display
configuration as necessary.
(
"Numeric Data/Waveform" P13-4)
1
Press the key for "ECG1" to "ECG12".
The "Size" screen will be displayed.
When the display layout is "12-Lead", the waveform size can be set
differently for limb leads and chest leads.
2
Select the waveform size for displaying and printing.
[Auto]: ECG amplitude will be automatically adjusted to 10mm.
The automatic adjustment is effective only when the [Auto] key is
pressed.
3
Use the
/
keys to adjust the baseline position.
If the waveform is difficult to see due to ECG amplitude, set the baseline position to 0mV. The baseline position
for the waveform display and printing will be adjusted. When the display layout is set to "12-Lead", the baseline
position cannot be changed.
Waveform Size
x1/4
x1/2
x1
x2
x4
Voltage (10mm)
4mV
2mV
1mV
500μV
250μV
3
2
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......