14
3.1
Level calibration
1.
If the difference of H1 and H2 is within 5mm-10mm,we may adjust
accuracy as following:
(1) Press ON/OFF,the indicator light on,it enter into self-leveling.
(2) Press G 3 seconds, X will twist on the LED, it enter into X direction
accuracy adjustment.
(3) Press +/-,laser beam will move step by step, check the laser beam
at Tilt mode, till the difference of H1 and H2 within 5mm,and press
G 3 seconds, the Tilt icon is disappear on the LED, the adjustment
is recorded.
(4) Press G 3 seconds, X is twist on the LED,then,press the X/Y switch
key is twist, it enters into Y direction adjustment.
(5) Press +/-,laser beam will move step by step, check the laser beam
at Tilt mode, till the difference of H1 and H2 within 5mm,and press
G 3 seconds, the Tilt icon is disappear on the LED, the adjustment
is recorded.