Rotary Laser   

 Service Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание FC101

Страница 1: ...Rotary Laser Service Manual ...

Страница 2: ...0 Structure 6 2 1Housing set 7 2 2Underpan set 8 2 3Pendulum 10 2 4PCB 13 2 5 Instrument Assembling 14 3 0 Instrument Inspect 15 3 1 Instrument level adjusting 15 4 0 Horizontal line Adjustment 17 4 1 Inspection 17 4 2 Adjustment 17 5 0 Detector 17 6 0 Malfunctions Eliminations 19 ...

Страница 3: ...emote control and charger FC101 It can assemble 4pcs rechargeable battery only have to loose the screw on the base And take the battery off 1 1 Basic function 1 Press the ON OFF it has a rotary laser beam form a horizontal Main boday Remote control FRD100detector and clamp FC101charger ...

Страница 4: ...yclable it will display the on LED It has 5 steps 0 60 120 300 600 r p m 3 Direction scanning recyclable the angle will be display on LED 5 steps from 0 10 45 90 180 4 Slope setting Can switch the laser unit between manual mode and self leveling mode There is a triangle icon on LED 5 left spining It adjusts the slope when in slope setting mode and it will be display on the LED It can control the l...

Страница 5: ...xis or Y axis to set the slope 8 Tilt mode When the instrument is set in Tilt mode when it leveling any movement or shock it out of leveling and indicate user the statement is changed it has TILT display on LED 2 0 Structure FRE207 2 1001Housing set FRE207 2 200 base set FRE207 100Pendulum set ...

Страница 6: ...lass 1 3 FRE207 110 Top rubber cover 1 4 FRE206 003 connecting board 1 14 FRE206 001 lamina Lucida 1 15 LED 1 16 GB T845 1985 screwST2 2 4 5 4 5 Menu sticker 1 6 FRE101 302 Shielding glass 1 1 7 FRE203 214 Gasket ring 1 Housing bombing illustration 1 2 3 4 7 5 8 11 9 10 6 12 13 15 16 14 ...

Страница 7: ...ing terminal 4 10 FRE101 315YB Receiver board 1 12 FRE206 002 Middle Housing 1 13 FRE203 206 Bottom rubber cover 1 11 GB T845 1985 Screw ST2 2 4 5 4 2 2 Housing set Speed up left spinning Direction scanning right spinning X Switch ON OFF Manual Illustration of Menu function Tilt Slope setting ...

Страница 8: ...8a Base 1 5 FRE203L 308 Press board 1 6 FRE203 204 Battery cover 1 7 FRE203 208 battery fixing frame 2 8 FRE203L 312 socket cover 1 9 FRE203N 007 battery seal ring 1 10 FRE203N 008 battery seal ring II 1 11 FRE203L 306 tight nut 1 12 FRE207 117 Bubble base 1 Bombing illustration of Underpan set 1 2 3 4 5 6 10 8 7 11 9 12 ...

Страница 9: ...E NAME Quantity FRE101 307 pentagonal prism 10 10 1 FRE102 410 optical wedge fixing base 1 FRE102 411 optical wedge base 1 1 FRE102 412 optical wedge 1 Pendulum set bombing illustration 1 4 5 2 3 8 9 10 6 7 11 13 14 15 16 17 12 ...

Страница 10: ...2 GB T80 2000 Screw M3 10 8 6 GB T96 1985 φ3 gasket 2 FRE203 105 Universal holder 1 GB T70 1 2000 Screw M3 15 4 7 GB T80 2000 Screw M2 5 3 2 FRE207 104 leveling frame 1 8 GB T818 2000 Screw M2 8 4 9 FLE441 107 Pin 2 10 FRE207 103 Fixing board 1 11 FRE203 211 Alarm base 1 P7 106 Alarm spring 1 11 GB T818 2000 Screw M2 5 6 2 FRE203 106 Universal ring 1 FLE441 107 Pin 2 GB T80 2000 Screw M2 5 3 2 12 ...

Страница 11: ...ward 1 16 1526 diode downward 1 Bubble base set 1 17 GB T80 2000 screw M3 8 6 2 5 Main PCB X motor socket Y motor socket Direction scanning socket Alarm socket scanning motor socket Upper laser socket Down dot socket Power socket Charger socket Sensor socket Keyboard socket Infrared receiver socket ...

Страница 12: ...12 2 6 Assembling Housing set Pendulum set Base set ...

Страница 13: ...wall or scaleplate as depicted below 2 Switch on the power then measure the h1 of X1 beam on the wall or scaleplate by using detector 3 Loose the screw of the tripod then turn around the instrument 180 o to measure the h2 of X2 beam on the wall or scaleplate D value between h1 and h2 ought to be less than 10mm 4 Inspect the Y beam with the same way Assembling bombing illustration X1 beam X2 beam X...

Страница 14: ...ep by step check the laser beam at Tilt mode till the difference of H1 and H2 within 5mm and press G 3 seconds the Tilt icon is disappear on the LED the adjustment is recorded 4 Press G 3 seconds X is twist on the LED then press the X Y switch key is twist it enters into Y direction adjustment 5 Press laser beam will move step by step check the laser beam at Tilt mode till the difference of H1 and...

Страница 15: ...ed adjusting mechanic as following As the above picture just X1 or X2 Y1 or Y2 under Self leveling statement till the difference is within 5mm Attention If the calibration is out of range You should better contact distributor for repairing X1 Y2 X2 Y1 B A C D ...

Страница 16: ...strument 3 Switch on the power and then measure the middle point of the laser beam on two walls or scaleplates There are hA hB and hA hB 4 1 hA hA 2 hB hB D value between 1 and 2 ought to be less than 4mm 4 2 Adjusting Vertical If D value between Δ 1and Δ 2 is more than 4mm adjust the instrument in the same way of level adjusting Please refer to the details of Y axis calibration in 3 0 and 3 1 hA ...

Страница 17: ...of the detector s circuit 1 The laser beam is received by silicon photocell After amplifying and judging the signal received the detector will send out command signal to control the display of LCD and the buzzer States of buzzer upward bias alarm downward bias alarm accurate orientated alarm prolonged sound 2 Normal fault and repair 1 Buzzer doesn t work Symptons Buzzer off when press key LCD illu...

Страница 18: ...method a Open the batteries cover then reset the batteries b Open the shell then insert the socket c Fault of circuit board send back to manufacturer 6 0 Malfunctions Eliminations 6 1 Failures in starting up 1 Symptoms After pressing the Key ON OFF power indicator can t light and also the instrument doesn t start up 2 Causes a Poor contact of keys on the panel or panel plug b Poor contact of batte...

Страница 19: ...p after all your effort above 6 2 No spinning of laser head 1 Symptoms The laser beam is being emitted but the laser head doesn t rotate after instrument is power on 2 Causes a Center to center distance of gear is not suitable gearing is abnormal b Troubles of scanning motor c Laser head has touched the optical coupler d Troubles of PCB 3 Eliminations a Take off the shell to readjust the center to...

Страница 20: ...20 ...