Table 11: Transducer model:
C8x
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
<
1
A
aprt
>1
Global maximum index value
1.2
—
(a)
—
2.0
(b)
As
sociated
acoustic
paramete
r
p
r.3
(MPa)
2.59
W
0
(mW)
—
#
36.0
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
1.1
1.10
d
eq
(Z
sp
)
(cm)
0.28
F
c
(MHz)
4.79
—
#
—
4.79
#
Dim of A
aprt
X (cm)
—
#
—
1.12
#
Y (cm)
—
#
—
0.40
#
Other information
PD
(μsec)
1.131
PRF
(Hz)
1008
p
r
@PII
max
(MPa)
3.10
d
eq
@Pll
max
(cm)
0.28
Focal Length
FL
x
(cm)
—
#
—
#
FL
y
(cm)
—
#
—
#
I
PA.3
@MI
max
(W/cm
2
)
296
Operating
contr
o
l
con
d
ition
s Control 1: Exam type
Pro
Pro
Control 2: Sample volume size
1 mm
1 mm
Control 3: Sample volume position
Zone 5
Zone 5
Control 4: PRF
1008
3125
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.