Cleaning and disinfecting
111
10 Rinse
the transducer
3 separate times
using the following procedure:
a
Rinse the transducer in clean, running water according to the disinfectant manufacturer’s instructions
(at least 1 minute).
b
Ensure that the connector and at least 12-18 inches (31-46 cm) of the cable from the connector remain
dry.
c
Repeat this step until the transducer has been rinsed 3 separate times.
11 Dry
the transducer with a sterile, lint-free cloth.
12 Dispose
of the disinfectant according to the manufacturer’s guidelines.
13 Examine
the transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local
representative.
Clean and disinfect system and transducer to a low level
(non-critical uses)
Use the following procedure to clean and disinfect the ultrasound system and transducer
if it has not come
into contact with blood, broken skin, mucosal membranes, or bodily fluids
.
WARNING
If the system or transducer has come into contact with any of the following, use the
high-level cleaning and disinfection procedure. See Option A,
system and transducer to a high level (semi-critical uses)”
Blood
Broken skin
Mucosal membranes
Bodily fluids
B
Option
Содержание SonoSite Edge
Страница 1: ...Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
Страница 24: ...18 Compatible high level disinfectants compatible transducers...
Страница 46: ...40 Compatible high level disinfectants transducers...
Страница 68: ...62 Compatible high level disinfectants transducers...
Страница 88: ...82 Compatible high level disinfectants transducers...
Страница 110: ...104 Compatible high level disinfectants transducers...
Страница 150: ...144 Compatible high level disinfectants transducers...
Страница 151: ......
Страница 152: ...P22043 01 P22043 01...