108
Cleaning and disinfecting
Clean and disinfect system and transducer to a high level
(semi-critical uses)
Use this procedure to clean and high-level disinfect the ultrasound system and transducer
whenever it
has
come into contact with blood, broken skin, mucosal membranes, or bodily fluids
.
Follow the manufacturer's instructions when using cleaners and disinfectants. The cleaners and disinfectants
listed in the procedure are chemically compatible with the system and transducers. Confirm that the cleaners
and disinfectants are appropriate for your facility’s use.
Cleaning and disinfecting the system and transducer (semi-critical)
To clean and disinfect the system and transducer
1 Turn off
the system by pressing the
Power button.
2 Unplug
the power cord from the outlet.
3
Remove the disposable transducer sheath, if applicable.
4 Disconnect
the transducer from the system. Temporarily place it where it will not cross-contaminate clean
equipment or surfaces while you clean the ultrasound console.
WARNINGS
To avoid electrical shock, before cleaning, disconnect the system from the power
supply.
Wear the appropriate personal protective equipment (PPE) recommended by the
chemical manufacturer, such as eyewear and gloves.
Cautions
Do not skip any steps or abbreviate the cleaning and disinfecting process in any
way.
Do not spray cleaners or disinfectants directly on the system surfaces or on
system and transducer connectors. Doing so may cause solution to leak into the
system, damaging it and voiding the warranty.
Do not attempt to disinfect a transducer or transducer cable using a method or
chemical that is not included here. This can damage the transducer and void the
warranty.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage the System and transducer and void the warranty. Follow the
disinfectant manufacturer’s recommendations for solutions strengths.
Note
You must clean and disinfect both the ultrasound system and the transducer after
each use, but only the transducer can be disinfected to a high level.
Option A
Содержание SonoSite Edge
Страница 1: ...Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
Страница 24: ...18 Compatible high level disinfectants compatible transducers...
Страница 46: ...40 Compatible high level disinfectants transducers...
Страница 68: ...62 Compatible high level disinfectants transducers...
Страница 88: ...82 Compatible high level disinfectants transducers...
Страница 110: ...104 Compatible high level disinfectants transducers...
Страница 150: ...144 Compatible high level disinfectants transducers...
Страница 151: ......
Страница 152: ...P22043 01 P22043 01...