897N102548
Frequent messages and countermeasures
is displayed. / Images are not transferred.
Not transferred image data remain in the
Image Reader. Do not switch off the power.
CAUTION
►Cause
- The read image data could not be transferred to the
Console (PC).
- The power supply of the Console (PC) is switched off.
- No Exposure menu has been registered.
The cassette has not been inserted correctly.
►Cause
The cassette has not been inserted correctly into the
Image Reader.
Do not pull out the cassette forcibly.
A part of the device may become broken.
If the IP protrudes out of the cassette pulled
out, return the IP into the cassette and close
the lid of the cassette.
CAUTION
Image reading is not completed.
►Cause
A cassette has been ejected without reading the images
exposed to IP.
Images remain on the IP.
Repeat image reading.
CAUTION
2
1
Stop the buzzer.
Press the (Stand-by Switch) to stop the buzzer.
2
Confirm the Console.
- Confirm correct operation of the Console (PC).
- When the main power of the Console (PC) is off, turn
on the main power to start up the Console (PC).
- Confirm that an Exposure menu has been registered.
3
Confirm the cable connection.
- Confirm that the network cable connecting the Image
Reader and the HUB has not been disconnected.
- Confirm that the network cable connecting the Console
(PC) and the HUB has not been disconnected.
- Sometimes the error may be corrected by reconnecting
the network cable.
- Confirm that the power cord for the HUB has not been
disconnected.
Countermeasures
★:チェックポイント
Check Point
Countermeasures
1
Stop the buzzer.
Press (Stand-by Switch) to stop the buzzer. Press
(Stand-by Switch) again to clear the error indication.
2
Remove the cassette slowly.
Confirm that the
(Cassette Removal Lamp) and the
(Cassette Ready Lamp) are blinking alternately,
press the (Stand-by Switch) again and then remove
the cassette.
3
Align the cassette to the right and insert it
slowly.
Confirm that the cassette is not upside down or switched
front/back.
Insert the cassette following the operation instruction
attached to the Image Reader.
1
Stop the buzzer.
Press (Stand-by Switch) to stop the buzzer. Press
(Stand-by Switch) again to clear the error indication.
2
Remove the cassette.
Confirm that the
(Cassette Removal Lamp) and the
(Cassette Ready Lamp) are blinking alternately,
press the (Stand-by Switch) again and then remove
the cassette.
3
Perform image reading.
Insert the removed cassette into the unit and perform
image reading again.
Countermeasures