Transanal Insertion
6.2.2
(1) Instruct the patient to take an appropriate position for the
insertion procedure.
(2) Unlock the bending portion by turning the up-down
locking lever and the left-right locking knob in the
direction of F until they stop.
[Note]
Another procedure is also available: you can insert the
endoscope by locking the bending portion only in the
left-right direction and unlocking it in the up-down
direction.
(3) Switch on the power to the processor and turn on the lamp.
(4) Remove air from the balloon on the distal end of the
endoscope and the balloon on the overtube.
Apply clean lubricant (Xylocaine jelly or the like) to the
insertion portion as required.
[Note]
Do not apply Xylocaine spray, olive oil or the like directly
to the insertion portion.
6-10
Chapter 6 Method of Use