Table 2

Guidance and manufacturer’s declaration - electromagnetic immunity

The CR-IR346CL/348CL is intended for use  in the electromagnetic environment specied below.

The customer or the user of the CR-IR346CL/348CL should assure that it is used in such an environment.

Immunity  test

IEC  60601  test  level

Compliance  level

Electromagnetic  environment  -  guidance

Electrostatic discharge (ESD)
IEC61000-4-2

±6kV contact

±8kV air 

±2kV contact
±4kV contact
±6kV contact

±2kV air 
±4kV air 
±8kV air 

Floors should be wood, concrete or ceramic tile. If oors are covered

with synthetic material, the relative humidity should be  at least 30%.

Electrical fast transient/burst
IEC61000-4-4

±2kV for power supply lines

±1kV for input/output lines

±2kV for power supply lines

±1kV for input/output lines

Mains power quality should be that of a typical commercial or hospital
environment.

Surge
IEC61000-4-5

±1kV differential mode

±2kV common mode

±0.5kV, ±1kV differential
mode

±0.5kV, ±1kV, ±2kV
common mode

Mains power quality should be that of a typical commercial or hospital
environment.

Voltage dips, short interruptions
and voltage variations of power
supply input lines
IEC61000-4-11

<5%

(>95% dip in

)

for 0.5 cycle

40%

(60% dip in

)

for 5 cycles

70%

(30% dip in

)

for 25 cycles

<5%

(>95% dip in

)

for 5 s

<5%

(>95% dip in

)

for 0.5 cycle

40%

(60% dip in

)

for 5 cycles

70%

(30% dip in

)

for 25 cycles

<5%

(>95% dip in

)

for 5 s

Mains power quality should be that of a typical commercial or hospital
environment. If the user of the  CR-IR346CL/348CL requires continued
operation during power mains interruptions, it  is recommended that the
CR-IR346CL/348CL be powered from an uninterruptible power supply or
a battery.

Power frequency (50/60Hz)

magnetic eld

IEC61000-4-8

3

 

A/m

3

 

A/m

Power frequency magnetic elds should be at levels characteristic of a

typical location in a typical commercial or hospital environment.

NOTE:

 is the a.c. mains voltage prior to application of the test level.

Cautions Relating to Electromagnetic Compatibility

Содержание CR-IR346CL

Страница 1: ...FCR XG 1 CR CL FCR XG 1 CR CL CR IR346 348CL CR IR346 348CL Service Manual Service Manual...

Страница 2: ......

Страница 3: ...ch items are dispatched or shipped to foreign countries and when service training is given for trainees from foreign countries Export Restrictions Applying to this Equipment Precautions in Using the S...

Страница 4: ......

Страница 5: ...erences to mammography related options Mammography functionality for this product is not commercially available in the US pending FDA Approval This functionality must not be activated on US models the...

Страница 6: ......

Страница 7: ...BLANK PAGE...

Страница 8: ......

Страница 9: ...he release of this document may differ from the descriptions given in this document In such an instance a Service Engineering Information will be issued as needed to furnish latest information about i...

Страница 10: ...herwise be sure to turn off the power switch before maintenance work and disconnect the power plug from the outlet Even if the power switch is OFF 100 to 240VAC differs according to machine speci cati...

Страница 11: ...acture label CR IR348CL Lite CR IR348CL Plus Identification label CL01002 EPS Identi cation labels 00010008 EPS Year of Manufacture and EFUP Environmentally Friendly Use Period label 2007 Sample year...

Страница 12: ...ification label CE label 00010011 EPS Identi cation labels CR IR348CL Lite CR IR348CLPlus 00010012 EPS CR IR348CL Mobile model CE labels CR IR348CL Lite CR IR348CLPlus 00010013 EPS CR IR348CL Mobile m...

Страница 13: ...ers and heat sources Precautions when working on internal parts of the control unit 1 Turn off the power of the PC CRT and other peripheral devices 2 Disconnect the power cord from the outlet to preve...

Страница 14: ...RF communications equipment can affect medical electrical equipment 3 Information of all cables and their maximum lengths and accessories that may affect EMC Name Speci cation Network cable Cat5e or...

Страница 15: ...mon mode 0 5kV 1kV differential mode 0 5kV 1kV 2kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variation...

Страница 16: ...y an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80...

Страница 17: ...aration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For tra...

Страница 18: ...CR IR346CL CR IR348CL RU CR IR346RU OS Operating System Windows 2000 XP Vista AP Application Software Name of the Target Product of This Manual Note that the target product name of this manual has be...

Страница 19: ...t When to Obtain and to Bring Out Personal Information Follow the operating procedures under the Private Information Protection Law and prescribed by your local FUJIFILM representative When you collec...

Страница 20: ...BLANK PAGE...

Отзывы: