8. APPENDIX

6SHFL¿FDWLRQ

  General Specifications 

Measuring object: 

Oxygen in noncombustible gas 

Measuring method: 

 

 

Directly insert type zirconia system

Measuring range: 

  0 to 2 ··· 50 vol%, two ranges, user 

configurable

 

(in 1 vol% O

steps)

Repeatability: 

 Within 

±0.5%FS

Linearity: 

 Within 

±2%FS

Response time:

  

Within 4 to 7 sec, for 90% (from calibra-
tion gas inlet)

Warmup time:

  

More than 10 min

Analog output:

  

4 to 20mA DC (allowable load resistance 
less than 500

:

) or 0 to 1V DC (output 

resistance more than 100

:

)

Power supply:

  

Rated voltage; 

100 to 120V AC (operating voltage 90 
to 132V AC)
200 to 240V AC (operating voltage 
190 to 264V AC)

 

Rated frequency; 50/60Hz

Power consumption:

 

 

Maximum 240VA (Detector: approx. 
200VA, Converter: approx. 40VA)

 

During normal operation 70VA (Detec-
tor: approx. 50VA, Converter: approx. 
20VA)

  Detector Specifications (ZFK)

Measured gas temperature:

 

Flow guide tube system; −10 to +600ºC       
(for general-use, corrosive gas)

 

Ejector system; −10 to +1500ºC (for 
high-temperature gas)

 

−10 to +800ºC (for general-use)

Measured gas pressure: 

 

−3 to +3kPa (−306 to +306mmH

2

O)

Flow guide tube:

   With or without blow-down nozzle

 

Flange; JIS5K 65A FF

   

(JIS5K-80AFF for high particulate gas)

 

Insertion length; 0.3, 0.5, 0.75, 1m

Ejector (general-use): 

   

Probe for guiding measured gas to 
detector 

   

Flange; JIS10K 65A RF

   

Insertion length; 0.5, 0.75, 1, 1.5m (ac-
cording to customer's specification)

Operating temperature: 

   

−10 to +60ºC for Primary detecting ele-
ment

   

−5 to +100ºC for ejector section

   

125ºC or less at detector flange surface 
with power applied

  

 When sample gas temperature is 
lower than 150°C and the outside 
temperature is lower than 0°C, cover 
the flow guide tube flange and the 
detector (the part that contact outside 
air) with thermal insulating material to 
prevent dew condensation.

Storage temperature:

   

Sensing element: −20 to +70ºC

   

Ejector: −10 to +100ºC

Structure: 

 

Dust/rain-proof structure(IEC IP66 
equivalent)

Filter: 

 

Alumina(filtering accuracy 50μm) and 
quartz paper

Main materials of gas-contacting parts:

   

Detector; Zirconia, SUS316, platinum

   

Flow guide tube; SUS304 or SUS316

   

Ejector (general use); SUS316, SUS304

   

Ejector; (for high temperature) SiC, 
SUS316, SUS304

Calibration gas inlet:

 

   

I

6mm tube join, 

I

1/4-inch tube join, or 

ball valbe (as specified)

Reference air inlet (option):

 

   

I

6mm tube join or 

I

1/4-inch tube join (as 

specified)

Detector mounting:

 

   

Horizontal plane ±45º, ambient sur-
rounding air should be clean.

Outer dimensions:

 (L × max. dia.) 210mm × 100mm (de-

tector)

Mass (approx.) {weight}: 

 

  

Detector; 

1.6kg

   

Ejector; 15kg (insertion length 1m)

   

Flow guide tube (general-use, 1m); 5kg

Finish color: 

 

Silver and SUS metallic color 

Ejector air inlet flow rate:

 

   

5 to 10 L/min

Calibration gas flow:

 

   

1.5 to 2 L/min

Blowdown air inlet pressure:

 

   

200 to 300kPa {2 to 3 kgf/cm

2

}

Ejector exhaust gas processing:

 

   

Into furnace, returned to flue

Heater temperature drop alarm output (ejector):

   

Alarm output when below 100

 

ºC  Me-

chanical thermostat

   

N.O. (1a) contact, 200V AC, 2A

All manuals and user guides at all-guides.com

Содержание ZFK8 Series

Страница 1: ...Instruction Manual THE DIRECT INSERTION TYPE ZIRCONIA OXYGEN ANALYZER DETECTOR TYPE ZFK8 INZ TN5ZFK8d E All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 2: ...J WKLV PDQXDO LW VKRXOG DOZD V EH NHSW LQ D SODFH ZKLFK DOORZV WKH SHUVRQ ZKR RSHUDWHV LW WR refer to any time as requirfed GXH FRQVLGHUDWLRQ VKRXOG EH JLYHQ VR WKDW WKLV LQVWUXFWLRQ PDQXDO LV GHOLYHU...

Страница 3: ...LRQV VHW IRUWK LQ LQ VWUXFWLRQ PDQXDO 7KH XVH DW D SODFH QRW FRQIRUPLQJ WR WKH LQVWDOODWLRQ FRQGL WLRQV PD UHVXOW LQ DQ HOHFWULF VKRFN D UH DQG PDOIXQFWLRQ KHQ WKLV SURGXFW LV PRXQWHG RQ D IXUQDFH ZKL...

Страница 4: ...WLQJ D EXUQ HIRUH FOHDQLQKJ WKH RZ JXLGH WXEH WXUQ RII WKH PDLQ SRZHU DQG FRRO WKH WXEH GRZQ IXOO There is a fear of getting a burn RQ W XVH DQ RWKHU UHQHZDO SDUW WKDQ WKRVH GHVLJQDWHG E WKH PDNHU 2WK...

Страница 5: ...G 3LSLQJ 3LSLQJ RI FDOLEUDWLRQ JDV 3LSLQJ RI UHIHUHQFH JDV LQOHW 3LSLQJ IRU EORZ GRZQ DLU 3LSLQJ GUDZLQJ 8 LULQJ HIRUH ZLULQJ LULQJ WR HDFK WHUPLQDO 0RXQWLQJ RI FRQGXLW 2SHUDWLRQ DQG 6WRS 6WDUW RI RSH...

Страница 6: ...URO VLJQDO SURFHVVLQJ RXWSXW GLVSOD DQG H WHUQDO WUDQVPLVVLRQ 7KH GHWHFWRU DQG FRQYHUWHU DUH FRQQHFWHG ZLWK D FDEOH 1 INTRODUCTION High oxygen partial pressure side PH Low oxygen partial pressure side...

Страница 7: ...6HH WHP Thermo sticker FFHVVRULHV VK 6HH WHP 6 HUDPLF OWHU FFHVVRULHV VK 6HH WHP ORZ JXLGH WXEH 6SHFL FDWLRQ JRRGV V RUGHUHG 6HH WHP DQG 8 HDW LQVXODWLQJ FRYHU 6SHFL FDWLRQ JRRGV V RUGHUHG 6HH WHP 5H...

Страница 8: ...K DOORZV GRLQJ GDLO FKHFN DQG ZLULQJ ZRUN 3ODFH ZKHUH WKHUH LV OLWWOH YLEUDWLRQ GXVW DQG KXPLGLW 3ODFH ZKHUH SHULSKHUDO DLU HQYLURQPHQW LV QRQ FRUURVLYH 3ODFH ZKHUH WKHUH DUH QR HOHFWULF DSSOLDQFHV SU...

Страница 9: ...H WR RU OHVV UHJDUGOHVV RI WKH PHDVXUHG JDV WHPSHUDWXUH How to check RPSOHWH WKH PRXQWLQJ SLSLQJ DQG ZLULQJ RI WKH GHWHFWRU PDNH VXUH WKH FRORU RI WKHUPR VWLFNHU SXW RQ WKH GHWHFWRU DQJH LV QRW FKDQJL...

Страница 10: ...KH SDUWLWLRQ SODWH LQVLGH WKH RZ JXLGH WXEH WR WKH JDV RZ DQG PRXQW WKH WXEH VR WKDW WKH WRQJXH WXUQV WR D GRZQ VWHDP GLUHFWLRQ UHODWLYH WR JDV RZ Mating flange not included in scope of supply Gas flo...

Страница 11: ...4 Mounting of heat insulating cover RU XVLQJ DW D FROG GLVWULF XVH D KHDW LQVXODWLQJ FRYHU RU WKH SDUW 1R IRU DGGLWLRQDO SURFXUHPHQW RI WKH KHDW LQVXODWLQJ FRYHU UHIHU WR WHP 3XW 0 PRXQWLQJ VFUHZ RI W...

Страница 12: ...6HW RQH RI WKH WZR UHIHUHQFH JDV SRUWV DV LQOHW DQG WKH RWKHU DV RXWOHW QVWDOO SLSLQJ VR WKDW GXVW RU ZDWHU GRHV QRW HQWHU WKH RXWOHW 3LSLQJ IRU EORZ GRZQ DLU RQQHFW WKH EORZ DLU LQOHW ORFDWLRQV RI EO...

Страница 13: ...act input Alarm contact output Maintenance contact output Calibrating gas contact output Note 1 Standard gas or instrumentation air can be used in place of span gas 2 Instrument quality air or bottled...

Страница 14: ...ermocouple for combustion control Contact input Alarm contact output Maintenance contact output Calibrating gas contact output Note 1 Standard gas or instrumentation air can be used in place of span g...

Страница 15: ...ut Heater HEATER Thermocouple for combustion control Contact input Alarm contact output Maintenance contact output Calibrating gas contact output Option 2 Ref air to Detector Flowmeter Flowmeter Valve...

Страница 16: ...noid valve not supplied Reduction valve ZBD6 Reduction valve ZBD6 Span gas Standard gas for zero point ZBM Joint Ejector system AC power supply O2 input Thermocouple input Analog output 4 to 20mA DC o...

Страница 17: ...URFHHGLQJ ZLWK WKH ZLULQJ ZRUN EH VXUH WR WXUQ RII WKH PDLQ SRZHU VXSSO 2WKHUZLVH WKHUH is a fear of getting an electric shock V WKH ZLULQJ PDWHULDO XVH D SURSHU RQH FRQIRUPLQJ WR WKH UDWLQJ RI DSSDUD...

Страница 18: ...U PRUH LQ UDWLQJ RU 5 WKHUPRFRXSOH 6SHFL HG LQ 6 HTXLYDOHQW WR 5 PP 6 5HFRPPHQGHG ZLUH URG DW Composition For heater Element output compensation conductor 1RPLQDO VHFWLRQDO DUHD PP 1XPEHU RI FRPSRVHG...

Страница 19: ...RI WKH PDLQ ERG LW WKH SDFNLQJ KROGHU QXW DQG SDFNLQJ RQWR WKH FRQGXLW DQG W WKH HQG IDFH RI FRQGXLW LQWR D JURRYH RI WKH EXVKLQJ QVHUW WKH EXVKLQJ WWHG WR WKH FRQGXLW LQWR 0 FRXSOLQJ RI WKH PDLQ ERG...

Страница 20: ...IHUHQFH JDV VXSSO LW EHIRUH WKH FDOLEUDWLRQ RZ UDWH WR PLQ 5 2 Stop of operation I WKH SRZHU LV WXUQHG RQ LQ LQ D VWDWH RI GHZ FRQGHQVDWLRQ LW OHDGV WR WKH IDLOXUH RI GHWHFWRU 6WRS WKH RSHUD WLRQ IROO...

Страница 21: ...RI WKH FRQYHUWHUV 0 50 DQG 5 FKHFN RQFH D ZHHN DV D URXJK VWDQGDUG FDOLEUDWLRQ JDV RZ UDWH WR PLQ KHQ WKH FRQYHUWHUV 0 50 DQG 5 DUH QRW XVHG 5XQQLQJ WKH HUR DQG VSDQ FDOLEUDWLRQ JDVHV PDNH FDOLEUDWLR...

Страница 22: ...XQLW DQG R ULQJ IURP WKH GHWHFWRU WWDFK WKH VXSSOLHG 2 ULQJ RI WKH UHSODFHPHQW VHQVRU XQLW WR WKH JURRYH IRU WKH 2 ULQJ DW WKH GHWHFWRU WWDFK WKH UHSODFHPHQW VHQVRU XQLW WR WKH GHWHFWRU WWDFK WKH VHQV...

Страница 23: ...QJ ZLWK WKH XVH RI D VFUXEELQJ brush 5HPRYH GXVW VWLFNLQJ WR WKH LQVLGH RI WKH RZ JXLGH WXEH E XVLQJ D PHWDOOLF URG RU VFUHZGULYHU OHDQ VR WKDW WXEH LV WKURXJK DW OHDVW DERXW SDUW RI WKH ZKROH LQWHULR...

Страница 24: ...WRU UHIHU WR WKH VWDQGDUG RXWSXW WDEOH EHORZ 6WDQGDUG RXWSXW WDEOH 5HIHUHQFH 2 JHQ concentration 9RO Detector output 8QLW P9 2 JHQ concentration 9RO Detector output 8QLW P9 2 JHQ concentration 9RO Det...

Страница 25: ...SLHFHV DUH UHTXLUHG IRU WKH VWDQGDUG JDV LQOHW RXWOHW 5HIHUHQFH JDV LQOHW IRU LQFK WXEH GGLWLRQDOO procurement part 3 7 3 DVNHW LQFOXGHG SLHFHV DUH UHTXLUHG IRU WKH VWDQGDUG JDV LQOHW RXWOHW 5H HFWRU...

Страница 26: ...HSODFH FDEOH UURU LQ ZLULQJ KHFN ZLULQJ LUH FRUUHFWO RZ VXSSO YROWDJH KHFN LI VXSSO SRZHU LV DV VSHFL HG 6XSSO FRUUHFW SRZHU LVFRQQHFWLRQ RI WKHUPR couple KHFN FRQWLQXLW KHFN LI UHVLVWDQFH DFURVV WHUP...

Страница 27: ...erature 10 to 60 C for Primary detecting ele ment 5 to 100 C for ejector section 125 C or less at detector ange surface with power applied When sample gas temperature is lower than 150 C and the outsi...

Страница 28: ...on in 4 digits Contact output signal 1 Contact speci cation 6 points 1a 250V AC 3A or 30V DC 3A 2 Contact function s 5NDER MAINTENANCE s 5NDER BLOWDOWN OTE s 3PAN CALIBRATION GAS VALVE s ERO CALIBRATI...

Страница 29: ...particulate with 1000mm cover Others Protection cover Without With Power supply 100 to 120VAC 50 60Hz 200 to 240VAC 50 60Hz Cal gas inlet For 6mm tube SUS For 1 4 inch tube SUS Ball valve 0 Y 0 5 A 3...

Страница 30: ...e detector side Both sides Types For R thermocouple K R YA YB YC YD YE YF YG YH YJ YK YL YM AA BB CC DD 0 1 2 Note5 For connection between detector and converter the conduit to be used should be rainp...

Страница 31: ...e Or 1 4 inch tube Ground wire Screw M4 80 67 6 0 Approx 62 Approx 132 Approx 130 5 79 26 Approx 30 125 5 85 5HSODFHPHQW VHQVRU XQLW 3 1 6 T H S 4 2 5 T H S made in japan TK4J4664R0 4 3 5 For detector...

Страница 32: ...e 11th Z F K 8 R 5 5 B 3 5 7 1 67 6 M5 detector side ORZ JXLGH WXEH ZLWK EORZ GRZQ QR OH 5 7 1 18 Approx L Approx 20 67 60 5 Gas inlet 8 15 MTG holes 6 M5 detector mounting side 67 130 155 Z F K 8 R 5...

Страница 33: ...or prevention of drain 67 29 5 1 4 5 1 8 0 4 Rc1 4 with plug Flange JIS 5K80A FF ZFK MTG position 50 30 185 Approx L Gas flow Gas outlet Packing Tube 50A SCH40 3 5 7 1 Z F K 8 R 5 6 D 3 L m 5 7 1 4 5...

Страница 34: ...ge is approx 150mm Do not block the gas outlet 175 Approx L 10 Approx L Gas flow Gas outlet Packing 38 Protection tube 65A SCH40 Tube 50A SCH40 3 5 7 1 Z F K 8 R 5 6 E 3 L m 5 7 1 7 1 9 0 11 4 13 6 L...

Страница 35: ...ors Planning Dept 1 Fuji machi Hino city Tokyo 191 8502 Japan http www fujielectric com Phone 81 42 514 8930 Fax 81 42 583 8275 http www fujielectric com products instruments All manuals and user guid...