![Fuji Electric FSC-1 Скачать руководство пользователя страница 194](http://html1.mh-extra.com/html/fuji-electric/fsc-1/fsc-1_instruction-manual_2341884194.webp)
INF-TN3FSC-E
185
)ORZYHORFLW\SUR¿OHGDWD¿OH
$¿OHLVJHQHUDWHGLQWKH&69IRUPDWLQGLFDWHGEHORZDQGLWFRQWDLQVGDWDIRUXSWRRQHKRXU
7KHIROORZLQJWDEOHLQGLFDWHVFRQWHQWVRIOLQHVRIURZVLQFDVHWKH¿OHLVRSHQHGZLWK([FHO
Table 16-3
&RQWHQWVRIÀRZYHORFLW\GLVWULEXWLRQ¿OH
Line
Row
Contents
Line 2
A
Describes "<Measurement result>".
B and subsequent
Describes "Channel number".
C and subsequent
Channel number
126 to 1, 1 to 126 in case of connection of sensor U
1 to 126, 126 to 1 in case of connection of sensor D
1 to 126, 126 to 1 to IT row in case of connection of sensor U/D
IU
Describes "<F.RATE/VEL.VALUE>".
IV
Describes "<RAS>".
Line 3
B
Describes "<Start/End Channel>".
C and subsequent
Describes 1 in start/end channels of channel number in line 2, and
describes 0 in others.
Line 4
A
Describes "<Measurement time>".
C
Describes acquisition date/hour in the "2007/10/29 10:19:44" format.
Line 5 and
subsequent
A
'HVFULEHV9HORFLW\3UR¿OH7KHPHDVXUHPHQWGDWD!
C to IU
Fluid velocity that corresponds to channel number in line 2.
IU
)ORZUDWHÀXLGYHORFLW\DWDFTXLVLWLRQGDWHKRXU
IV
5$6LVLQGLFDWHGE\ELWELQDU\QXPEHUVZLWKDGGHGDW¿UVW
See "16.5.4 Regarding RAS ".
$ÀRZYHORFLW\GLVWULEXWLRQ¿OHFDQEHGLVSOD\HGXVLQJÀRZYHORFLW\GLVWULEXWLRQGHPRQVWUDWH
IXQFWLRQRI3&ORDGHUVRIWZDUH
16.5.4 Regarding RAS
([SUHVVHVWKHVWDWXVRIWKHPHDVXUHPHQWVFUHHQE\ELWELQDU\QXPEHUV
Bit31
0
3
7
11
15
19
23
27
"0" expresses normal.
"1" expresses abnormal.
□□□□□□□□□□□□□□□□□□□□□□□□□□□□□□□□
E1: Device error 1
(error in backup memory)
E1: Device error 2
(error in measurement circuit)
E2: Window scanning
E2: No received signal
E2: Received signal error
E2: Calculation error 1
(error in fluid acoustic velocity)
E2: Calculation error 2
(receiving signal overflow)
E3: Measuring range error
E3: Frequency calculation error
E3: Success rate
E4: Range over
E6: Printer fail
6HH&RQWHQWVRIHUURUVLQVWDWXVGLVSOD\IRUHUURUFRQWHQWV