XM310K Calibration Guide
© 2020 Flanders Scientific, Inc.
9
Once the DDC reset is complete select NEXT to continue to the 1D LUT calibration page.
If you are
calibrating in Dynamic (modulated backlight) Luminance mode skip this step,
as it is only sup-
ported in static backlight modes. For Dynamic Luminance Mode ensure that your baseline EOTF set-
tings on the monitor correspond to the color space you are targeting.
Static Luminance Modes Only: From the 1D LUT calibration page click on the AutoCal symbol and
confirm that you have the LUT ID position set to the correct memory position. CalMAN will then begin
the 1D LUT AutoCal process. It is normal for the screen to flash during this process as the LUTs are
updated in real time. The probe will only read between flashes so any lines, flashes, etc. that occur
during a LUT update will not negatively impact calibration. Once this process is complete select NEXT
to continue to the 3D LUT calibration step.
All Luminance Modes: From the 3D LUT calibration page select the AutoCal symbol. Confirm your LUT
settings once again to ensure that the 3D LUT will be saved to the correct memory position. We also
suggest using 0.5 pattern delay.
The XM310K is a very linear device so a Lightning LUT may be more than sufficient to generate excel-
lent calibration results. We suggest trying Lightning LUT first before trying longer calibration processes,
but time based and point based selections will also work well. Once the calibration process is complete
the 3D LUT will be automatically saved to the display.
Additional Considerations
Please verify you are running the latest firmware on your monitor before starting calibration.
When verifying HDR calibration results on the XM310K it is advisable to use an L30 or L32 size pattern.
Questions? E-mail: [email protected] or Call: +1.678.835.4934