background image

This system has pre-installed Tee fittings for measuring Post 
Membrane and Post DI filter. 
The Red lead (in) should be installed into the tee located at the back 
of the system. The Blue lead (out) should be installed into the tee 
located on the outlet of the DI filter (right side). When inserting the 
probes, you will need to push them in firmly until they pass the 
internal o-ring and ‘lock’ into place. Try to orient the probes so that 
the pins are perpendicular to the water flow (shown in photo). 

You can mount the TDS monitor wherever it is convenient using the supplied adhesive tape. 

III. Usage 

1.  Press the “POWER” button. 
2.  To display the TDS level of the feed (tap) water, slide the switch to the IN side. To display the TDS 

level of the product (filtered) water, slide the switch to the OUT side. 

3.  The displayed TDS will be most accurate after approximately 10 seconds. 
4.  Determining filter effectiveness depends on your particular system. For an RO system, for example, 

compare the IN water TDS levels with the OUT water TDS. 

5.  If the “x10” icon appears, this means the TDS level is above 999 ppm. Therefore, multiply the 

reading by 10. For example, if the display shows 143 ppm with the ‘x10’ icon, the actual TDS level is 
1430 ppm. (If the ‘x10’ icon does not appear, the reading on the display is the actual TDS level.) For 
most drinking water, you will not see the ‘x10’ icon. 

6.  Turn off the unit. It will automatically shut off after 10 minutes. 

IV. Calibration 

Your monitor was factory calibrated to 342 ppm. This level is suitable for most tap water/filtered water 
applications, so it is ready to use out of the box. However, you may need to re-calibrate based on your 
needs, as well as from time-to-time to ensure best results. To calibrate: 

1.  Purchase a certified calibration solution that is correct for your needs. The calibration solution 

should be NaCl. HM Digital’s 342 ppm NaCl is recommended. 

2.  Disconnect both T-fittings from their tubes. Do not remove the sensor from the T! Ensure the 

orientation of the sensor to the fitting is correct, as in illustration #1. Shake any water out. 

3.  For better accuracy, calibrate to a flowing solution. If this is not possible, you can calibrate to a still 

solution. Turn on the monitor and place one T-fitting (with the sensors in it) into the calibration 
solution. You will get a reading. Ensure the fitting is completely filled with solution and there are no 
air bubbles. This step is critical for proper calibration. 

4.  If the reading on the monitor (for the sensor in the solution) does not match the solution, adjust the 

reading up or down by gently turning the plastic orange screw on the rear of the unit clockwise or 
counter clockwise to raise or lower the reading. 

5.  There is one calibration screw to calibrate both the IN and OUT sensors simultaneously. You only 

need to calibrate one sensor, and it does not matter which. 

6.  If calibrating to a still (not flowing) solution, calibrate to 3% above the level of the calibration 

solution. This will accommodate for the lack of flowing water, which the monitor is programmed for. 
For example, if the calibration solution is 342 ppm, adjust the screws until it reads 352 ppm. If you 
are calibrating to a flowing solution, calibrate to the level of the solution. 

7.  Your monitor is now calibrated. There is no need to do anything else 

V. Changing Batteries 

If the display is fading or the readings are not correct, you may need to change the batteries. 

1.  Unscrew the three metal screws (not the orange plastic screw) on the rear of the unit and remove 

the back panel. 

2.  Remove the batteries. 
3.  Replace both batteries with two fresh batteries (model 357A). Ensure the polarity is correct. 
4.  Close the back panel and replace the screws. You will not need to recalibrate. 

Содержание GT1-39RODI

Страница 1: ...k Adaptor Supplied as Standard 3 IV Connecting Tubing Using Compression Fittings 3 System Installation Start Up 4 I Pre Flushing 4 II Tubing Connections 4 III Initial Start Up 4 IV Using the System 5...

Страница 2: ...rger molecules from the water carrying them to waste Reverse Osmosis membranes do not have a litre Capacity for filter replacements due to the variables that can make the filter life vary significantl...

Страница 3: ...the Kitchen Diverter Valve but this time you firstly turn off the cold water to your washing machine lay a towel down for water leakage Using either your hands or a pair of gripping tools slowly unscr...

Страница 4: ...ure everything looks sound there are no kinks damaged fittings or tubing that is not plugged in Connect the inlet filter to the inlet side of the membrane screw cap end You should have 2 lengths of tu...

Страница 5: ...the desired level 5 If the TDS level is still too high open the flush valve for another 30 seconds then close and check VI After Use Storage The system can then be stored away in a cabinet if you have...

Страница 6: ...ution that is correct for your needs The calibration solution should be NaCl HM Digital s 342 ppm NaCl is recommended 2 Disconnect both T fittings from their tubes Do not remove the sensor from the T...

Страница 7: ...DS reading exceeds your desired level The membrane itself has a largely varying life time It may last anywhere between 1 and 4 years depending on usage water quality and age This can also be monitored...

Страница 8: ...side if it is damaged it may need replacing this is a common size washer and can be purchased from most hardware plumbing stores 2 If the diverter valve is loose or you can easily swivel the fitting w...

Страница 9: ...specified amount 3 If you have good pressure and average water quality it is possible there is a small blockage somewhere in the system Try to follow the water flow along the system disconnecting 1 t...

Страница 10: ...there is a strong foul smell or organic taste to the water it is possible that there is some form of contamination Contact us straight away so we can rectify or diagnose the problem if there is one p...

Страница 11: ...lter 6 Months GT4 1C Clear Refillable Cannister Post Filter 12 24 Months GT5 12S MB 115 Tulsion De Ionising Resin as Required GT13 37 ALP 100GPD Reverse Osmosis Membrane 1 4 years TDS Testing Required...

Страница 12: ...A plumber is not legally required for this type of installation unless a new tap or water outlet is required The system is compliant to WaterMark AS NZS 3497 and therefore is compliant with permanent...

Страница 13: ...s extended warranty covers the below parts of the system GT8 31 Reverse Osmosis Housing GT8 0S Twin O ring White Housing GT1 39 Models Only GT8 38S Twin O ring Clear Housing GT1 39 Models Only The fol...

Отзывы: