40
FONIX FP35 Quick Reference Guide
4.5.1 Viewing the SPL Display
The information on the SPL screen is similar to the information in the other real-
ear measurement screens. However, in addition to the measurement curves and the
SPL-converted target, the SPL Screen also displays SPL-converted thresholds and
uncomfortable levels, allowing you to directly compare the measurement curves to
the audiometric information.
X denotes left thresholds
O denotes right thresholds
*
denotes the target
U denotes the uncomfortable levels.
Figure 4.5.1—SPL display
4.5.2 Running an SPL Test
This section assumes you have already entered the client’s audiometric data in the
Audiogram Entry Screen, and that you have generated a target. See Section 4.3.3 and
4.3.4 for more details.
1. Place the probe tube into the client’s ear following the instructions provided in
Section 4.2.4.
2. Insert the aid into the client’s ear, being careful not to drag the probe tube farther
into the ear canal (See Figure 4.2.4C). Do not turn the aid on.
3. Instruct the client to remain still when the signal is on.
4. Press [F5] and [START/STOP] to level.
5. Have the client turn the aid on and set the gain control to the normal use level.
Содержание FONIX FP35
Страница 1: ...Quick Reference Guide 2008 Frye Electronics Inc All rights reserved P N 010 0021 04 FONIX FP35 ...
Страница 2: ......
Страница 18: ...14 FONIX FP35 Quick Reference Guide Figure 2 2 5 IM distortion measurement ...
Страница 21: ...3 Automated Test Sequences 17 Figure 3 1 2A ANSI test completed Figure 3 1 2B ANSI 96 test I O Screen ...
Страница 24: ...20 FONIX FP35 Quick Reference Guide Figure 3 2 2B IEC test results I O Screen ...
Страница 54: ......