6.3 Calibration plate
Two different calibration plates are available for calibrating the ANALY-
SETTE 28 ImageSizer. Order these plates according to the lenses you are
using. These calibration plates are used as follows:
1.
Remove the chute lid.
2.
Move the feeder as far back as possible. (See
‘Installing the feeder and funnel’ on page 34)
3.
The calibration plate has the designations "camera" and "flash".
The "camera" side must point to the left in the direction of the lens
and the camera, the "flash" side to the right towards the feeder
and flash.