8
Cleaning
8.1 Cleaning the devices
DANGER!
Mains voltage!
– Before beginning with cleaning work, disconnect
the mains plug and protect the device against being
unintentionally switched back on!
– Do not allow any liquids to flow into the device.
– Indicate cleaning work with warning signs.
– Put safety equipment back into operation after
cleaning work.
The devices can be cleaned using a moist cloth or the supplied microfibre
cloth.
8.2 Cleaning the hoses
You can clean stubborn soiling in the hoses of the wet dispersion unit by
adding the included cleaning powder.
Proceed as follows for cleaning the hoses:
1.
Empty the sample bath.
2.
Remove the hoses on the measuring cell drawer and connect both
hoses with the supplied connector. This creates a new liquid circu-
lation system.
3.
Fill the sample bath with liquid.
4.
Set the pump power to maximum and add 10 ml of cleaning
powder to the sample bath.
5.
Let the pump run for about 2 minutes. Then rinse the system 2 - 3
times to rinse out all the cleaning powder.
Cleaning
- 37 -