Using the device
6.1 Wet dispersion unit
6.1.1 Fritsch test powder F500
You can check the precision and accuracy of your device
by measuring the Fritsch test powder at least once a
week, depending on how often you use the device. This
will indicate contamination or malfunctions in your
system.
Measure the supplied Fritsch test powder F500 with the wet dispersion
unit as follows:
1.
Switch the measuring unit on the main switch to I (On). If you are
using an ultrasonic box, switch it on as well by the main switch.
2.
Open the liquid inlet.
3.
Start the MaScontrol software.
4.
Create a new root folder.
5.
An "Empty Folder" must be created in the newly created root
folder.
6.
Double-click the symbol in front of the newly created folder. A
measuring results folder opens.
Using the device
- 31 -