2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 505
Yes
Pump calibrates okay! See procedure number
(page 505).
No Pump does NOT calibrate! See procedure number
(page 507).
Figure 86 –
UF Pump
Adjustment
UF- 2.1.0 UF PUMP CALIBRATES OKAY / ISOLATE UF CIRCUIT IN DIALYSIS
a) Plug into concentrate and return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)
b) From the Home screen, allow [Temperature] and [Conductivity] to become normal!
c) ENSURING the external flow indicator’s ‘bob’ is rising and falling, reset ALL alarms and select the
screen’s
[Test & Options]
button.
d) Select
[Both Tests
] then press ‘CONFIRM’. Allow the ‘Conductivity’
test to pass.
e) Remove the Fluid Sample Connector to observe UF output below.
f) Call debug screen 0 to see the UF Pump’s ‘dot’ and TMP (Figure right).
g) WHEN
“Test: (Get Neg TMP)”
is up AND TMP is less than 240, watch
the Fluid Sample Connector. When the UF Pump’s ‘dot’ blinks between
white and blue, about every eight (8) seconds, there should be strong
pulses through the Sample Connector that squirt out at least six (6)
feet into the room.
Yes
Strong pulses! The UF Pump is okay! See procedure number
(page 505).
No Weak or no pulses! Proceed to
page 507
, procedure number
UF- 2.2.0 STRONG PULSES / VERIFY PRESSURE TESTS
a). TMP should increase to more than 240 mmHg causing the “Test: Neg Flow On” banner to appear.
Allow both (Negative and Positive) pressure test to complete.
b). Do BOTH tests pass?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...