2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 404
SECTION 7 - COND REF FAILURE OR COND OFFSET FAILURE
CR- 1.0.0 ISOLATE DISTRIBUTION BOARD
a)
Turn the machine OFF!
b) Figure right, ENSURE distribution board
position “X4, PH-PR” is COMPLETELY
VACANT!
NOTE:
If the CBE board is accidentily
plugged into position “PH-PR” it may destroy
the Power Logic board!
c) Turn the machine ON.
d) If the “Cond Ref Failure” OR “Cond Offset Failure” reoccurs see procedure number
CR- 1.0.1 VERIFY COND FAILURE
Has a Sensor and / or a Functional Board been recently been ‘swapped’ in?
Yes
If ‘swapping’ for troubleshooting purposes,
other than
“Cond Ref Failure” OR “Cond Offset
Failure” refer to
OPERATING MODES
(page
Error! Bookmark not defined.
) to place the
machine into
T and C Mode
.
No See parts a THROUGH e below:
a)
IMPORTANT!
Before continuing to part b, perform
(page 6) and return
here if a problem is NOT located.
b) To avoid pulling cables loose GENTLY open the card cage.
c) Figure below, at the rear of the machine, ENSURE the 24V POWER harness has remained
plugged in!
d) Spread the card cage side panels open then gently drop the front panel down.
Parts e and f next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...