![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 74](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626074.webp)
74
7.3
Error Codes
Error Code
Description
Recommended Actions
01
Motor rotation
Check correct functioning of the motor.
Check the ribbon cable connection between the optical switch
and the CPU board (J5 connector). See figure 2.8, page 16.
Check the correct functioning of the motor rotation optical switch.
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
02
Displacement
Check the displacement sensor calibration.
Run test 6 Displacement. See section 6.2.7, page 45.
Check the displacement sensor connection to CPU board
(J2 connector). See figure 2.8, page 16.
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
10
RAM
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
11
Flash memory (deleting or
programming problem)
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
12
Resources
Carry out a complete configuration for the pump using the
Agilia Partner maintenance software.
If the problem cannot be resolved:
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
13
Secondary Crc16
microcontroller - version is
incorrect
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
14
Secondary microcontroller
restarting - Watchdog
problem
Replace the Power supply board. See section 8.5, page 88.
15
Secondary microcontroller
communication
Check the ribbon cable connection between Power supply board
(J1 connector) and CPU board (J6 connector).
See figure 2.8, page 16.
See figure 2.10, page 20.
Replace the Power supply board. See section 8.5, page 88.
16
Disconnected battery
Check the battery connection to power supply board
(J3 connector). See section 2.1.5, page 20.
Check the battery charging voltage. Run test 12.
See section 6.2.13, page 53.
Replace the battery. See section 8.1, page 79.
17
Battery charge
Check the battery charging voltage. Run test 12.
See section 6.2.13, page 53.
Replace the battery. See section 8.1, page 79.
18
AC power presence
Remove the battery door. See section 8.1, page 79.
Check the status of the AC power fuse.
Replace the Power supply board. See section 8.5, page 88.
19
Battery temperature
Check the battery status. Run test 8.
See section 6.2.9, page 50.
Replace the battery. See section 8.1, page 79.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...