![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 53](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626053.webp)
53
6.2.13
Test 12: Battery Life
6.2.14
Test 13: Battery Test
Test Description
This test provides information on the current battery charge percentage, and displays results of
the last battery discharge test (last run of "Test 13: Battery test").
The battery capacity is acceptable.
If the charge is low, connect the pump to an AC power supply to charge the battery.
If the displayed "Alarm" is less than 01h30 (01:30), replace the battery.
Required Tools and Equipment
No specific tools or equipment are required
Prerequisites
Run test 13: "Battery test" on the pump before launching this test.
See section 6.2.14, page 53.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 12: Battery life
.
3.
Press
enter
.
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Charge
: Battery charge current rate (in %)
Auto Off
: Date of last automatic cut off (value saved during last battery
test)
Pre alarm
: Duration between transition to battery and pre alarm (value
saved during last battery test)
Alarm
: Duration between transition to battery and alarm (value saved
during last battery test)
Cut off
: Duration from alarm to automatic pump power off (value saved
during last battery test)
5.
Press
exit
to return to the
Tests
menu.
Test Description
During this test, a battery discharge is performed. Pre alarm, alarm and cut off times are
determined and registered so that it is possible to evaluate if the battery works correctly.
At the end of a battery test, the results are recorded and can be checked by running
"Test 12: Battery life". Then the pump powers off.
The battery capacity is acceptable.
Replace the battery.
Required Tools and Equipment
No specific tools or equipment are required
Prerequisites
The pump is disconnected from the AC power supply.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...