Troubleshooting and Messages
7.1
Alarm Messages and Information Signals
Displayed Message
Priority
Problem / Resolution
ERxx(yyyy) !!!
High (!!!)
Technical alarm. See section 7.3, page 74.
Battery alarm !!!
High (!!!)
The battery is discharged.
The pump will power OFF automatically within 5 minutes.
Connect the pump to a power supply immediately.
Very low battery !!!
High (!!!)
Very low battery.
Connect the pump to a power supply and allow time to
charge.
End of infusion !!!
High (!!!)
The infusion is completed (simple rate).
Occlusion alarm !!!
High (!!!)
The pressure in the infusion line has reached the threshold
level.
Check whether the infusion line is occluded. If necessary,
readjust the pressure threshold.
Syringe installation !!!
High (!!!)
The syringe is not installed correctly (plunger driver, syringe
barrel clasp or flange detection).
Check the syringe installation.
End of volume limit !!!
High (!!!)
The volume limit is reached.
End of volume/time !!!
High (!!!)
The VTBI is completed.
End of dose/time !!!
High (!!!)
Plunger head alarm !!!
High (!!!)
The plunger head is missing or incorrectly inserted.
Check the syringe installation.
Disengagement mechanism !!!
High (!!!)
The Disengagement mechanism is open after selection of a
syringe.
Check the syringe installation.
Battery pre-alarm !!
Medium (!!)
Low battery.
Connect the pump to a power supply.
Occlusion pre-alarm !!
Medium (!!)
In-line pressure has reached the following value:
25 mmHg / 2.5 kPa / 0.5 PSI below the programmed
threshold (from 50 to 250 mmHg).
50 mmHg / 5 kPa / 1 PSI below the programmed threshold
(over 250 mmHg).
Check the infusion line.
Set the correct pressure threshold.
Near end of infusion !!
Medium (!!)
The time remaining is less than the defined time duration
(adjustable between 1 and 30 minutes), and the remaining
volume of fluid in the syringe has dropped to less than 10% of
the syringe capacity.
Near end of volume limit !!
Medium (!!)
The time remaining is less than the defined time duration
(adjustable between 1 minute and 30 minutes), and the
remaining VTBI until the volume limit has dropped to less than
10% of the syringe capacity.
Near end of volume/time !!
Medium (!!)
The time remaining is less than the defined time duration
(adjustable between 1 and 30 minutes), and the remaining
VTBI has dropped to less than 10% of the syringe capacity.
Near end of dose/time !!
Medium (!!)
Check settings !!
Medium (!!)
The flow rate (or dose) has been modified using the keys, but
has not been confirmed.
Check the flow rate (or dose) and press
OK
to confirm.
Waiting settings !!
Medium (!!)
A value must be entered.
Enter a value and press
OK
to confirm.
Waiting start !!
Medium (!!)
The infusion settings have been entered, but have not been
confirmed with
start
.
Check the infusion settings, and press
start
to start the
infusion.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...