Chapter 9: Troubleshooting
Fresenius Kabi CompoSeal Universal OP-EN 6/07.13
23
9
Troubleshooting
CompoSeal Universal
Check the sealing head for proper seating
Clean the sealing head.
Check the fuses and the power cable.
Sparking
An alarm is sounded,
spark
LED is illuminated.
Sparking may occur if the tubing to be sealed is humid or if sealing is
interrupted by a sealing error.
Re-seal the tubing.
Sealing time exceeded
An alarm is sounded, the
time limit
LED is illuminated.
The sealing time has been exceeded.
Re-seal the tubing
Note
For all other problems, except for the causes described below, or in the
event of repeated problems, contact your nearest authorized
Composeal Service to ensure fast and proper diagnostics and
correction of the problem.