background image

 

90PGA 

Section 2 

Dual Channel Programmable 

Remote Operation 

Amplifier Card 

 

   

6

  

 

1784 Chessie Lane, Ottawa, IL 61350 • Tel: 800/252-7074, 815/434-7800 • FAX: 815/434-8176

 

e-mail: [email protected] 

 Web Address: http://www.freqdev.com 

during driver/libraries installation. Do not reboot until the end of the install procedure, where a 
single popup window prompts you for a restart. The installation process should not be 
interrupted with a reboot, otherwise installation must be re-started from the beginning.  

 

After installation and reboot, verify that the files reside in the selected target PC directory. An 
FDI icon should appear on you desktop. Note that verification of installed files after each step on 
the hard drive is optional. 

 

For serial remote operation: With the 90IP unit turned ON and the RS232 cable connected to 
COM1 or 2, RUN the NI-MAX 2.2.0 by clicking on the NI icon on the desktop, this updates the 
configurations. Wait for the data to be updated, indicated by no MAX or hard drive activity. Quit 
the MAX application window. 

 

Before initiating the remote mode for the 90IP, verify that it is powered up, connected to the PC 
Via the RS232 cable and that the display is not in remote mode.  Initiate the 90IP Instrument 
Control from the FDI desktop icon: Select RS232, and COM1 or 2 on the 90IP GUI.  Click on 
Scan and this operation will produce: ASRL<COM>::INSTR in the “Instruments” display window.  
Click on the items as they appear in the windows until the 90PGA GUI appears. Verify that 
remote control of the PGA card is functional (One can actually hear relays clicking for certain 
button changes in I/O).  Note: When finished, remember to use the Quit button to exit the 
application.  After quitting a remote session, press the ENTER button on the 90IP to return to 
local mode or turn the power button off and on, to reset the instrument. 

 

For GPIB remote operation: Connect a GPIB cable from the instrument to the PC, and turn on 
the 90IP as above. 

 

Initiate the 90IP Instrument Control from the desktop icon: Select 

GPIB

 then click on the Scan 

button:  GPIB0::1::INSTR should appear in the “Instruments” display window.  Click on items as 
they appear in the windows until the 90PGA GUI appears. Verify that control of the PGA card is 
functional. Use the Quit button to exit application. 

 
Initiating

 “SETUP”

 will toggle between install and uninstall. To uninstall all the 90IP-related software from 

the target PC, you may use the same disks for uninstalling the 

CDIP-01

 and 

CDFC-01

 software. This is 

accomplished by initiating the CD setups from each folder, again.  Each time this is done, the software 
toggles between install and uninstall. However, even though the NI libraries auto setup, they do not 
uninstall automatically. The NI libraries must be uninstalled separately using Add/Remove Programs in 
Control Panel. The list is as follows, NI IVI engine 1.83, NI-MAX 2.2.0, NI-488.2 1.70, NI VISA 2.6 Run-time 
Engine, and NI-VISA Server 1.0.0f6.  
 
See Appendix A for 

Windows 2000, Windows XP After Installation Considerations.

 

 

 

Содержание 90PGA

Страница 1: ...Slot Card Dual Channel Programmable Amplifier Card 1784 Chessie Lane Ottawa IL 61350 Tel 800 252 7074 815 434 7800 FAX 815 434 8176 e mail sales freqdev com Web Address http www freqdev com USER S MA...

Страница 2: ...1 1 Start Up 2 1 2 90PGA Operation Instructions 2 1 3 Front Panel Command Structure 3 Section 2 Remote Operation 2 1 Directory Descriptions 4 2 2 Installation and Set up 5 2 3 Slot Card Command Refere...

Страница 3: ...s 1 Upon power up a cursor appears under the default function value or mode Note the cursor is always initially displayed on the LSB of any numerical value 2 Turning the FUNCTION knob scrolls the curs...

Страница 4: ...r differential types for the input signal MODE AC AC coupled input SNGL Single Ended DC DC coupled input DIFF Differential Note Output is always single ended 5 GAIN Gain selection is performed by sett...

Страница 5: ...function cards available to date These directories install the code and the files used in developing the function card GUI s for custom development Follow the process instructions to install any utili...

Страница 6: ...2 to provide devices access and to configure hardware and software The CDFC 01 90IP Function Cards Installation Disk is provided for custom development and to install the application software needed...

Страница 7: ...n the items as they appear in the windows until the 90PGA GUI appears Verify that remote control of the PGA card is functional One can actually hear relays clicking for certain button changes in I O N...

Страница 8: ...ument controller set up Graphical User Interface GUI will appear 5 Select GPIB or RS232 For RS232 select the COM port 1 or 2 location where the RS232 interface card is installed in your PC 6 CLICK on...

Страница 9: ...eveloper need not worry about adding the command terminator to commands using GPIB Example You may need to drop the n character from the string you are using as a command No command processing occurs...

Страница 10: ...de of an active card channel to DC Set DC coupled mode AC Sets the coupling mode of an active card channel to AC Set AC coupled mode SNGL Sets the input mode of an active card channel to single ended...

Страница 11: ...84 Chessie Lane Ottawa IL 61350 Tel 800 252 7074 815 434 7800 FAX 815 434 8176 e mail sales freqdev com Web Address http www freqdev com 2 4 Main Soft Front Panel The components of the main panel are...

Страница 12: ...lection Sets the input mode for each channel in the slot card Select Single Ended for single ended input operation or Differential for differential input mode 4 Gain Control This control sets the gain...

Страница 13: ...lines suggesting a way to allow any desired user to acquire the necessary privileges to run the FD90IP Application Software Perform the following steps in order to grant the necessary privileges to th...

Страница 14: ...ight click on the 90IP Instrument Control folder and select Properties from the popup menu Click the Security tab on the 90IP Instrument Control window Select the Users group and allow full control by...

Страница 15: ...800 FAX 815 434 8176 e mail sales freqdev com Web Address http www freqdev com As a User 1 Double click on the MAX icon to run it 2 Refresh the configuration tree by clicking on F5 while the focus is...

Отзывы: