ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
44 / 78
4.
Service and Repair
4.1.
Service
Regular maintenances of the equipment are not essential, but we suggest an annual service
by the manufacturer to check all system settings and the operating ability. A vision control of
the whole system is recommended every six month.
Electrical wiring should be controlled by accreted firms during local mandatory intervals.
Avoid exposing the equipment to heat, dirt and other environmental influences. This will
elongate the lifetime of the equipment.
4.1.1.
Checking the system pressures
To check the different pressures of the measurement system use the panel with all pressure
information as shown in Figure 23.
Figure 23:
system pressures overview
The inert gas and air pressure can be changed by pulling out and turning the valves shown in
Warning
: It is not recommended to change the air pressure! For proper functionality the
system needs an air pressure of 5 bar.
Open the small panel on the
left side