background image

A control solution test should be performed when you are not sure of 
your results and want to confirm that your meter and test strips are 
working properly.

IMPORTANT:

Use only MediSense (low, medium or high) Glucose and Ketone Control Solutions 
with the meter. 
• Control solution results should fall within the control solution range printed on 

the test strip instructions for use. 

• Check that the LOT number printed on the test strip foil packet and instructions 

for use match. 

• Do not use control solution past its expiration date. Discard control solution 3 

months after opening or on the expiration date printed on the bottle, whichever 
comes first. (Example: open April 15, discard July 15; write the discard date on 
the side of the bottle.) 

• The control solution range is a target range for control solution only, not for 

blood glucose levels. 

• Replace the cap securely on the bottle immediately after use. 
• Do not add water or other liquid to control solution. 
• Contact Customer Care (1-800-527-3339) for information on how to obtain 

control solutions.

11

MediSense Glucose and Ketone Control Solution

Performing a Control Solution Test

1.  Open the foil test strip packet 

at the notch and 

tear down to remove the test strip.

2.  Insert the test strip 

until the meter turns on.

Notes:

• Be sure to check that your meter screen is working 

properly each time you turn your meter on. If you see 
any white segments in the black off screen, or any 
black segments in the white test screen, there may be 
a problem with the meter. (See Section 6, Getting to 
Know Your Meter, for more details.)

• The meter turns off after 3 minutes of inactivity. 

Remove and reinsert the unused test strip to restart 
the meter.

   Abbott Diabet

es C

ar

e L

td

.

Range R

oad

W

itney

, O

xon, O

X29 0Y

L

M

ade in UK

.

Blood Gluc

ose

Sensor Elec

trode

Abbott Diabet

es C

ar

e L

td

.

Range R

oad

W

itney

, O

xon, O

X29 0Y

L

M

ade in UK

.

4˚C

The    blinks, indicating the meter is ready for you to 
apply control solution to the test strip.

3. 

Press and hold the down arrow for 3 seconds to 

mark the test as a control solution test

. The 

  appears. The meter is now ready for you to apply 

control solution to the test strip.

IMPORTANT:

 The test result will be saved to 

memory as a blood result if not marked as 
a control solution test. This may affect your 
blood glucose averages.

4.  Apply control solution to the test strip.

Shake the control solution bottle to mix the solution. 
Apply a drop of control solution to the white area at 
the end of the test strip in the area shown. The control 
solution is drawn into the test strip.

5.  Hold the control solution to the test strip until:

• You see 3 short lines on the meter screen. This 

means you have applied enough control solution 
and the meter is reading the control solution.

Notes:

•  You will see a 5-second countdown. 
• Do 

not

 remove the test strip from the meter  during 

the countdown.

•  If the countdown does not start, remove and 

discard the used test strip, turn off the meter and 
try again with a new strip.

Example:

6.  View the result.

The test is complete (example shown) when the result 
appears on the meter screen. The result is stored in 
memory as a control solution result.
Compare the control solution result to the range 
printed on the blood glucose test strip instructions for 
use. The result should fall within the range.

Out of Range Control Solution Results:

• Repeat the test if control solution results are outside 

the range printed on the test strip instructions for 
use.

• Stop using the meter if control solution results are 

consistently outside the range printed on the test 
strip instructions for use. Contact Customer Care.

12

Transferring Meter Data to a Computer

Transferring meter data to a computer requires a compatible data management system. You 
will also need a micro USB cable to connect your meter’s USB port to your computer (available 
separately). For more information, contact Customer Care.

WARNING:

 To avoid the possibility of electric shock, never perform a blood glucose test 

while the meter is connected to the computer.

14

Error Messages

16

Troubleshooting

What It Means

What To Do

1.  Test strip 

is inserted 

in the 

strip 

port and 

nothing 

happens.

Test strip is not inserted properly or fully into the meter

1.

  With the contact bars (3 black lines) facing up, insert the test strip into the meter until it stops. 

This turns on the meter.

2.

  If the meter still does not turn on, contact Customer Care.

No batteries are installed; Batteries installed incorrectly

Refer to Section 15,

 Taking Care of Your Meter and Lancing Device

, on how to properly install 

batteries.

Dead batteries

Change batteries. Reset date and time, if needed.

Meter may be plugged into a computer (   appears on meter screen)

Unplug the meter from the computer.

Problem with the test strip

Try a new test strip.

Problem with the meter

Contact Customer Care.

2.  The test 

does not 

start after 

applying 

the blood 

sample.

Blood sample is too small

1.

 Repeat the test using a new test strip.

2.

 If the test still does not start, contact Customer Care.

Sample applied after meter turns off

1.

 Review the testing instructions.

2.

 Repeat the test using a new test strip.

3.

 If the test still does not start, contact Customer Care.

Problem with meter or test strip

1.

 Repeat the test using a new test strip.

2.

 If the test still does not start, contact Customer Care.

17

Meter Specifications

Message

What It Means

What To Do

The temperature is too hot 
or too cold for the meter to 
work properly

1. Move the meter and test strips to 

a location where the temperature 
is within the test strip operating 
range. (See test strip instructions 
for use for the appropriate range.)

2. Wait for the meter and test strips to 

adjust to the new temperature.

3. Repeat the test using a new test 

strip.

4. If the error reappears, contact 

Customer Care.

Meter error

1. Turn off the meter.
2. Repeat the test using a new test 

strip.

3. If the error reappears, contact 

Customer Care.

Message

What It Means

What To Do

Blood drop is too small
or
Incorrect test procedure
or
There may be a problem 
with the test strip

1. Review the testing instructions.
2. Repeat the test using a new test 

strip.

3. If the error reappears, contact 

Customer Care.

The blood glucose level may 
be too high to be read by 
the system
or
There may be a problem 
with the test strip

1. Repeat the test using a new test 

strip.

2. If the error reappears, contact 

Customer Care.

Message

What It Means

What To Do

Blood was applied to the 
test strip too soon

1. Review the testing instructions.
2. Repeat the test using a new test 

strip.

3. If the error reappears, contact 

Customer Care.

Meter error

1. Check that you are using the correct 

strip for this meter. (See test strip 
instructions for use to verify your 
strip is compatible with this meter.)

2. Repeat the test using a test strip for 

use with your meter.

3. If the error reappears, contact 

Customer Care.

15

Taking Care of Your Meter and Lancing Device

Changing Batteries

You will see this screen when your batteries are low.

Note: 

Your meter settings and logbook information will be saved 

when you change the batteries.

Note: 

When you no longer need the meter, remove the batteries and dispose of batteries and 

meter in compliance with your local government regulations. The batteries in this product should 

be removed and disposed of in accordance with local regulations for separate collection of spent 
batteries.

IMPORTANT:

 After you first see this warning, you can perform approximately 28 

tests before you need to change the batteries.

WARNING:

 Batteries should be kept away from small children. If swallowed, contact a 

healthcare professional immediately.

Step

Action

1. 

Turn meter over and slide open the battery door on 
the side as shown.

2. 

Remove the old batteries.

3. 

Install new batteries with (+) facing up.

Note: 

The meter uses 2 replaceable CR 2032 coin cell 

batteries.

4. 

Slide the door back into place until it clicks.

Note: 

The next time you turn your meter on, it may 

prompt you to reset the time and date. (Refer to Section 

7, Setting Up the Meter).

Cleaning and Disinfecting the Meter and Lancing Device

This section describes the cleaning and disinfection tasks for the FreeStyle Precision Neo 
Meter and the FreeStyle Lancing Device II. 

Cleaning and disinfecting your meter and lancing device is important to prevent the spread 
of infectious diseases. The meter has a mean use life of 5 years, which is 522 cleaning and 522 
disinfection cycles (2 cycles per week for 5 years). The lancing device has a mean use life of 2 
years, which is 210 cleaning and 210 disinfection cycles (2 cycles per week for 2 years). 

Turn off the meter prior to cleaning and disinfection. Users should wash hands thoroughly 
with soap and water after handling the meter, lancing device, or test strips.

If the meter is being operated by a second person who is providing testing assistance to the 
user, the meter and lancing device should be cleaned and disinfected prior to use by the 
second person.

Assemble cleaning and disinfection supplies.  

Clorox Healthcare Bleach Germicidal Wipes, EPA Reg. #67619-12. These disinfectant wipes 
contain a 0.55% Sodium Hypochlorite (NaOCl) solution and have been shown to be safe for 
use with the meter and lancing device. They may be purchased at major online retailers, such 
as Walmart.com, Amazon.com, and OfficeDepot.com.

·  Cleaning and disinfection should be performed using Clorox Healthcare Bleach Germicidal Wipes 

at least once per week. 

CAUTION: 

Do not place the meter in water or other liquids. Avoid getting dust, dirt, 

blood, control solution, water, bleach, or any other substance in the meter’s test strip 
port, USB port and battery compartment. 

·  The meter has a mean use life of 5 years and has been validated for 522 cleaning and 522 

disinfection cycles (the equivalent of 2 cycles per week for 5 years).

·  The lancing device has a mean use life of 2 years and has been validated for 210 cleaning and 

disinfection cycles (the equivalent of 2 cycles per week for 2 years).

·  Cleaning and disinfecting the meter is important to prevent the spread of infectious diseases. 

Cleaning and Disinfection

Cleaning is the physical removal of organic soil from the meter and lancing device surfaces. 
Keeping the meter and lancing device clean helps ensure that they are working properly and 
that no dirt gets in the devices. Cleaning allows for successful, subsequent disinfection.
Disinfection is a process that destroys pathogens, such as viruses and other microorganisms, on 
the meter and lancing device surfaces. Disinfecting the meter and lancing device helps ensure 
that no infection is passed on when you or others come in contact with the meter and lancing 
device. The meter and lancing device should be cleaned with a separate bleach wipe prior to 
each disinfection.
Clean and disinfect your meter at least once a week. Clean and disinfect your lancing device 
when visibly dirty (at least once a week).

Note: 

Squeeze bleach wipe to remove excess solution before wiping the meter or lancing 

device.

Assay method

Amperometry

Automatic shutoff

At least two minutes of inactivity

Battery life

Up to 3000 tests

Measurement 
range

For blood glucose testing 20 - 500 mg/dL

Memory

Up to 1000 events, including blood glucose and control solution 
results, and other meter information

Minimum computer 
requirements

System must only be used with EN60950-1 rated computers. Use a USB 
certified cable

Operating relative 
humidity

10% to 90% (non-condensing)

Operating 
temperature

59 °F to 104 °F (15 °C to 40 °C)

Power source

Two CR 2032 lithium (coin cell) batteries

Size

2.35 in (w) x 3.42 in (l) x 0.34 in (d)
5.97 cm (w) x 8.68 cm (l) x 0.87 cm (d)

Storage 
temperature

-4 °F to 140 °F (-20 °C to 60 °C)

System altitude

See test strip instructions for use

Weight

1.2 oz. to 1.3 oz. (33 g to 37 g) including batteries

Meter cleaning and 
disinfection

The meter has a mean use life of 5 years, which is 522 cleaning and 
522 disinfection cycles (2 cycles per week for 5 years).

Lancing device 
cleaning and 
disinfection

The lancing device has a mean use life of 2 years, which is 210 cleaning 
and 210 disinfection cycles (2 cycles per week for 2 years). 

Note: 

For test strip specifications, see test strip instructions for use.

Electromagnetic Compatibility (EMC): FreeStyle Precision Neo meter has been tested for both electrostatic discharge and radio 
frequency interference. Emissions are low and unlikely to interfere with other nearby electronic equipment. To limit radio 
frequency interference do not use the FreeStyle Precision Neo meter near cellular or cordless telephones, radio transmitters or 
other electrical or electronic equipment that are sources of electromagnetic radiation, as these may interfere with the proper 
operation of the meter. Avoid use of the device in very dry environments, as electrostatic discharges from synthetic materials 
(e.g., carpets) could cause damage. 

Distributed by:

Abbott Diabetes Care Inc.

1360 South Loop Road

Alameda, CA 94502 USA

Customer Care 1-800-527-3339

Monday through Friday, 8AM to 8PM 

Eastern Standard Time

FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various 

jurisdictions. Other trademarks are the property of their respective owners.

Patent: https://www.abbott.com/patents

©2016 Abbott  ART33078-003 Rev. A 09/16

Message

What It Means

What To Do

No coding required
or
Test strip may be damaged, 
used, or the meter does not 
recognize it

1. Check that you are using the correct 

test strip for this meter. (See test 
strip instructions for use to verify 
your strip is compatible with this 
meter.)

2. Repeat the test using a test strip for 

use with your meter.

3. If the error reappears, contact 

Customer Care.

Meter error

1. Turn off the meter.
2. Repeat the test using a new test 

strip.

3. If the error reappears, contact 

Customer Care.

18

References

References: 

1.  “FDA Public Health Notification: Use of Fingerstick Devices on More than One Person Poses Risk 

for Transmitting Bloodborne Pathogens: Initial Communication” (2010) http://www.fda.gov/
MedicalDevices/Safety/AlertsandNotices/ucm224025.htm

2.  “CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for 

Transmitting Bloodborne Pathogens” (2010) http://www.cdc.gov/injectionsafety/Fingerstick-
DevicesBGM.html

3.  American Diabetes Association. Standards of Medical Care in Diabetes—2013. Diabetes Care 

2013;6(Suppl.1)S11-S66.

Step

Action

1

For cleaning, wipe all outside surfaces of the meter and lancing device thoroughly 
with the bleach wipes until they are visibly clean.  Take care to avoid the strip port, 
USB port and battery compartment. Clean your meter at least once a week.
Clean your lancing device when visibly dirty (at least once a week).

2

Allow the meter and lancing device to air dry completely before performing a 
blood glucose test.

3

When finished, thoroughly wash your hands with soap and water. 

4

For disinfection, wipe all outside surfaces of the meter and lancing device 
thoroughly with the bleach wipe until they are wet in order to pre-clean the 
meter and lancing device. Take care to avoid the strip port, USB port and battery 
compartment.

Manufacturer: 

Abbott Diabetes Care Ltd.

Range Road

Witney, Oxon

OX29 0YL, UK

Step

Action

5

After pre-cleaning, use a second wipe to disinfect the meter and lancing device, 
using the same process. 

6

Allow the meter and lancing device surfaces to remain wet for 60 seconds (1 full 
minute). Allow the meter and lancing device to air dry completely before use. 
After 1 full minute, immediately dry with drying towel to remove any residual 
moisture. Disinfect your meter at least once a week.
Disinfect your lancing device when visibly dirty (at least once a week).  

7

Allow the meter and lancing device to air dry completely before performing a 
blood glucose test.

8

When finished, thoroughly wash your hands with soap and water.
Additional information about the risks for transmitting bloodborne pathogens 
to persons undergoing fingerstick procedures for blood sampling can be found. 
(See Section 18, References, for more details.)

IMPORTANT:

 If you require assistance or if you notice any signs of deterioration on the meter or lancing device (such as clouding or crazing on the display of the meter, corroding or eroding of the plastic housing, or cracking of plastic housing or display) or if the meter does not turn on or if the lancing device 

does not work after cleaning and disinfection, discontinue use of the meter or lancing device and contact Customer Care (1-800-527-3339).

13

View Logbook

Your meter logbook can store up to 1,000 events – including blood glucose and control 
solution results and other meter information.

Viewing Your Logbook Events

1. 

While on the Home screen, press       to view your 
logbook events.

Note: 

For today’s results, only the time will be displayed 

at the bottom of the screen. For previous days’ results, the 

date will alternate with the time. 

Examples:

Blood Glucose Result

2. 

Press    to scroll through to view your logbook events.

-

Control Solution Result

Note: 

Check mark in the box indicates a control solution 

test and not that the result was within range. 

Viewing Your Blood Glucose Averages

1. 

While on the Home screen, press      to open the 
logbook.

Note: 

For today’s results, only the time will be displayed at 

the bottom of the screen. For previous days results, the date 

will alternate with the time.

2. 

While in the logbook, press      any time to view your 
7-day average.   [ 

 ]

• Press    to view 14-day average.   [ 

 ]

• Press    to view 30-day average.   [ 

 ]

• Press      to return to your logbook events.

Notes:

• Averages do not include glucose control solution results.
• Control solution results not marked as control solution 

tests may cause averages to be inaccurate.

• 

   

appear on the meter screen when there are no 

current events or averages to view.

• 

   

blood glucose test results are included as 

20 mg/dL when calculating averages.

• 

   

blood glucose test results are included as 

500 mg/dL when calculating averages.

Press 

 or 

  to scroll 

through 7-, 14- and 30-day 
averages.

Your average

Number of tests 

in this average

Отзывы: