13. Maintenance and care
Check once a month that the eye shower is in working order. (Guidelines for Labora-
tories BGI/GUV-I 850-0)
As a preventative measure to reduce microbiological contamination, it is recom-
mended to replace the water in the fitting at shorter intervals by opening the valve.
Use suitable cleaning agents which are not harmful to the fitting in a proper manner
and rinse with water after use. High-pressure cleaners must not be used for cleaning.
14. Troubleshooting
If you cannot rectify a fault or if it is not in the list, please contact our customer
service department!
15. Spare parts
Fault
Cause
Remedy
Water does not flow
– Water supply interrupted
➯
Restore it
– Dirty filter
➯
Clean and replace
if necessary
Water volume too low
– Pressure reducer defective
➯
Replace it
– Dirty filter
➯
Clean it
Jet height too high
– Jet height set incorrectly
➯
Set it
– Flow pressure too high
➯
Assembly with
pressure reducer
Jet height too low
– Supply pressure too low
➯
Check it
– Pressure reducer set incorrectly
➯
Set it
– Flow pressure too low
➯
Assembly without
pressure reducer
Description
Order No.
1 Hand shower complete. . . . . . . . . 2000105863
2 Pressure reducer . . . . . . . . . . . . . 2000109525
3 Filter (packaging unit10x) . . . . . . . 2000109401
4 Dust cap with ball chain . . . . . . . . 2000105849
5 Spray head (packaging unit 4x) . . 2000104725
6 Non-return valve
(packaging unit 10x) . . . . . . . . 2000104744
7 Seal (packaging unit 10x) . . . . . . . 2000104668
8 Connection hose. . . . . . . . . . . . . . 2000105845
Maintenance set consisting of:
O-rings . . . . . . . . . . . . . . . . . . 2000105864